Page 518 -
P. 518

CHEMICALS AND CHEMICAL HANDLING           15.25


         vide a  safe  and  reliable  system.  All positive displacement  pumps  should  have  a  suitable
         pressure  relief valve installed in the discharge  piping  to prevent overpressurizing of pip-
         ing  due  to  inadvertently  closed valves or  other obstructions.  In  some  cases  these  valves
         are integral  with the pump.  Calibration  chambers  for calibrating the  discharge  from  me-
         tering pumps  should  be installed on  the  suction piping with valves and  fittings  to isolate
         it from the  chemical  supply  tank,  so that  measurements  can  be  made  with  the  pump  op-
         erating  against  normal  discharge  conditions.
           Back pressure  valves are used  to prevent gravity flow or siphoning  to a  low-pressure
         application point and to ensure consistent  operation.  A  typical set point for back pressure
         valves in this application would be at least 30 psig. Even when chemical feed pumps  nor-
         mally discharge to a high-pressure  application point,  an accidental loss  of pressure  in the
         system could result in an overdose by gravity flow or siphoning  action,  which  back pres-
         sure valves would mitigate.
           Figure  15.9  shows  a  typical  liquid  chemical  feed  system,  including  isolation  valves,
         strainers,  and a calibration chamber on the suction side and a series of safety valves (pres-
         sure relief, check, back pressure,  antisiphon)  along with isolation valves on the discharge
         side  of the  chemical  feed  pumps.  Vents  for  calibration  chambers  should  extend  above
         storage tanks  or whatever the  maximum  level of the  liquid is in the piping  system.
           Many positive displacement pumps  have integral check valves, and some are equipped
         with  combination  pressure  relief,  priming,  back  pressure,  and  antisiph0n  valves.  Even
         when a pump has its own combination of safety valves, the design should provide at least
         a redundant  antisiphon  valve to help protect the water  supply from an overdose of chem-
        ical,  should  a  system upset  occur.
           Low  pressures,  which  could  encourage  gases  to  come  out  of solution  and  form  bub-
        bles,  should  be  avoided.  This  requires  larger  suction  lines  that  are  as  short  as  possible.
         Suction piping  should have a positive static pressure  when possible.  A  standpipe  (or cal-
        ibration  column)  near the pump  suction  should  be considered  to divert bubbles  that may
        be present  in the  suction line.
           Positive displacement pumps typically require some type of pulsation dampener on the
         discharge  (and  sometimes suction) piping to reduce intermittent high pressures  and partly
        equalize the flow of liquid chemical to the application  point. Pulsation  dampeners  should
        be placed upstream  of back pressure  valves to work properly.  Even with pulsation  damp-
         ening,  the piping pressure  rating  should be  at least  150%  of average working pressure.
           Valves  for  isolation  should  be  used  often  on  chemical  feed  piping.  Even  long  lines
         with  no  attachments  or  equipment  should  be  equipped  with  isolation  valves for mainte-
         nance and cleaning. Unions should be installed around pumps  and valves to facilitate their
         removal.

        Eductors.  Eductors,  also referred to  as ejectors or jet pumps,  are useful  for introducing
         gases,  solutions,  or  slurries  into  a  diluted  water  solution  followed  by  application  to  the
         process  water.  A  stream  of water passing  through  a  venturi in the  eductor  creates  a  sub-
         atmospheric  pressure,  which  draws  the  chemical  into  the  eductor.  Because  eductors  are
         incapable  of flow rate  control, the  chemical must  be metered  in  some fashion  before en-
         tering the eductor.  The chemical introduced at the eductor can be monitored by a flowme-
         ter and  its  flow rate  controlled by  a  valve in response  to  a  reading  from  the  meter.  This
         practice should be used  only when  adequate  suction pressure  is available from the chem-
         ical supply  source,  to allow the control valve to accurately control the chemical  entering
         the  eductor.

         Centrifugal Pumps.  Centrifugal pumps  are used primarily in chemical feed systems  for
         chemical transfer  from bulk tanks  to day tanks.  Centrifugal pumps  for chemicals operate
         similarly to centrifugal pumps  for water.  Speed  variation  affects both  the discharge flow
   513   514   515   516   517   518   519   520   521   522   523