Page 39 - Glucose Monitoring Devices
P. 39

The process for premarket approval of SMBG devices     35




                  and conditions (e.g., ketoacidosis) and an evaluation of the stability of reagents and
                  materials, but these changes only minimally affect the analytical performance of
                  BGMs.
                  Precision
                  This term represents how closely and consistently repeated measurements per-
                  formed by SMBG on a given sample align with each other. According to the latest
                  ISO 15197 2013 guidelines, testing must demonstrate repeatability on at least five
                  venous blood samples with defined glucose concentrations in the hypoglycemic,
                  euglycemic, and hyperglycemic ranges. These guidelines also assess intermediate
                  measurement precision by requiring daily measurements over 10 days.

                  Interference evaluation
                  The 2013 update to ISO 15197 introduced the evaluation of performance in the
                  presence of various hematocrit ranges and other possible interfering substances
                  that might be found in the blood such as medications or physiologically occurring
                  substances. For hematocrit, the BGM must be tested in five different hematocrit
                  ranges at each of three defined glucose concentrations. Other interfering substances
                  must be tested for at least two defined glucose concentrations. In each of these
                  different testing environments, the ISO 15197:2013 standard allows for up to: 1)
                  10 mg/dL difference between the test sample and control sample for glucose
                  concentrations   100 mg/dL and 2) 10% difference for glucose concentrations
                  >100 mg/dL.

                  Accuracy
                  Accuracy refers to how closely the BGM system’s measurements align with refer-
                  ence values obtained by a gold standard technique. ISO 15197:2013 requires that
                  at least 95% of measurements fall within  15 mg/dL of the reference value for
                  blood glucose (BG) concentrations <100 mg/dL and within 15% at BG
                  concentrations   100 mg/dL. The previous 2003 version was less rigorous,
                  requiring  15 mg/dL of reference for BG concentrations <75 mg/dL and  20%
                  for concentrations   75 mg/dL.
                     In addition, the 2013 update introduced the requirement that at least 99% of
                  measurement results fall within the consensus error grid (CEG) zones A and B
                  [12]. The CEG has five zones, characterized by different clinical risks to the patient.
                  By emphasizing that 99% of results fall in zones A (no effect on clinical action) and
                  B (little or no effect on the clinical outcome), this addition reduces the maximum
                  amount of clinically unacceptable results to 1%, along with the maximum amount
                  of analytically unacceptable results to 5%.
                     Furthermore, the 2013 update increased the minimum number of test strips lots
                  being evaluated for accuracy from one to three, allowing for a more comprehensive
                  assessment to compensate for lot-to-lot variability in performance, which has been
                  shown to significantly affect the accuracy of a system [13].
   34   35   36   37   38   39   40   41   42   43   44