Page 440 - Handbook of Biomechatronics
P. 440
434 Graham Brooker
Fig. 15 A syringe insulin pump with adjustable dose rate.
by geared motors and ball screw mechanisms that depresses the plunger,
much as the original Mill Hill device did. They generally still deliver insulin
subcutaneously through a cannula. A drawing of a generic modern insulin
pump is shown in Fig. 16.
Insulin is measured in “units” where there are 100 units per mL, making
each unit 10μL. Basal rates are of the order of one unit per hour administered
every 3–10min, while bolus doses are typically several units. Cartridge vol-
umes are generally 200–300 units (2–3mL) and can last 4–5 days.
Because pumps are FDA regulated in the United States, their design and
construction must be precisely documented and their performance must
meet strict guidelines. In addition, the hardware must provide comprehen-
sive self-test and fault reporting capabilities. This requires significant
amounts of additional circuitry not directly involved in the pump process,
as can be seen in Fig. 17 (Mossman, 2010).
To ensure they are truly portable, insulin pumps should not weigh more
than 100g and be no larger than a mobile phone, yet should only require