Page 28 - Inorganic Mass Spectrometry - Fundamentals and Applications
P. 28

18                                                          S~it~

       nature; care must be exercised in these cases (although there is a NIST certified
       isotopic standard for lead).
            Certified standards or no, the same procedure is followed. The value of an
       isotopic calibration ratio is measured for the reference material and compared to
       the certified (or accepted) value. The correction factor necessary to adjust the
       measured value to the certified is calculated; it is then the bias correction for that
       ratio. Most often, a bias correction per mass is used.  Even though the actual
       variation of bias is not strictly linear, the limited mass range swept for a single
       element makes it a good approximation; any deviation fkom linear is insignificant
       in comparison to measurement uncertainties. Typical biases are a few tenths of a
       percent per mass when a single-fil~ent configuration is used and somewhat less
       for multifilament. Multifilament aqalyses are in! gener   sceptible to varia-
       tions in bias correction than single-filament, but they   no means immune,
            Bias corrections determined from analysis of s   are applidd to the
       samples under test. Use of such an average bias correction can be viewed only as
       an approximation to the truth; so many factors contribute to bias that it is impos-
       sible to control them all. For example, as previously stated, the work
       rhenium filament is determined by  which crystal face is involved:
       loading samples on filaments is through use of  single resin beads
       beads are 100-1200  pm in diameter, which is about the size of rhenium c~stal~ites
       in a polycrystalline filament [17]. Clearly the work function applicable to the
       analysis in question may or may not be that operative when ins~ment calibration
       was carried out. Another parameter difficult to control in real-world conditions is
       sample purity, which also affects bias. It is impossible to purify all samples to the
       same degree, and contaminants adversely affect ionization efficiency; low effi-
       ciency means higher filament temperatures, which in turn mean a different bias
       correction. These  are only two of  sundry variables that  can  affect ionization
       efficiency.
            In practice, the analyst monitors the bias correction through analysis of  a
       reference standard on a routine, often daily, basis. This value comes to be known
       very well and makes insignificant con~butions to overall precision. Even though
       it may not be truly applicable to the sample being analyzed, using it is far better
       tban applying no correction; it is the best that can be done in an imperfect world. A
       model of thermal fractionation on mass spectrometer filaments has been devel-
       oped by Habfast 1581.
            Fractionation is such a vexing problem that other means of  addressing it
       have been devised. One is total e~~a~stiu~, in which the entire sample is con-
       sumed; it has been successfully applied to uranium [59] and the rare earths 1601.
       The idea here is that, if all ions emitted from the sample are collected, they will be
       representative of  the  sample itself,  and  no  bias  correction  will  be  required.
       Because signal intensity varies rapidly, running to exhaustion can only be accom-
       plished using a multicollector mass spectrometer. It also requires a reasonable
   23   24   25   26   27   28   29   30   31   32   33