Page 78 -
P. 78

Ch13-I044963.fm  Page 61  Tuesday, August 1, 2006  12:49 PM
                      Page 61
                           Tuesday, August 1,2006
                                          12:49 PM
            Ch13-I044963.fm
                                                                                          61 61
                  Basic performance  evaluation of the forceps manipulator; working range, force, torque, and speed, was conducted
                  (TABLE  1, Figure 6). The  speed was measured  under two conditions: one was conducted with the  acceleration
                  time of 25[msec], the other was  1000[msec]. This was because the maximum speed of a stepper motor depended
                  on the acceleration time. We aimed to set the system control frequency  to be  10[Hz] (100[msec/cycle]),  and we
                  assumed that the acceleration time was at most quarter of cycle, 25[msec]. The speed of  1000[msec]  acceleration
                  was measured  as a case of enough long acceleration time. The working space was sector form whose radius was
                  340[mm] and whose vertex angel was  180[deg] in horizontal plane, and vertical depth was 360[mm] (Figure 6).



                  Bending  forceps
                  We  also measured  the  working  range,  positioning  accuracy, backlash, maximum  speed,  and torque  of bending
                  forceps (TABLE 2, Figure 6). Bending angle for "Bendingl" and "Bending2" was 304[deg] and 201 [deg] respec-
                  tively. The jaw  for  grasping  opened  up to 201 [deg]. Low positioning  accuracy was caused because of not  only
                  wire elongation and tension decrease, but also loose knot during the measurement. Output torque of "Bendingl",
                  "Jawl", and "Jaw2" were equivalent to the force;  1.9, 3.0, and 3.7[N] respectively. We found  abrasion and break
                  of the wire, showing that friction  occurred between the wire and path or pulley and that the transmission  efficiency
                  was reduced by the  friction.



                  In-vivo  experiments
                  We conducted in-vivo experiments on a swine to evaluate the system  in simulated clinical environment.  Setting-up
                  time was less than 30 minutes and we thought it clinically feasible. The slave robot was controlled using a master
                  manipulator by a surgeon (Mitsuishi  M,  et al. (2003)). Each DOF had enough working range, however, we  found
                  some problems. One problem was that; we inserted the surgical tool through trocar, an outer tube with air sealing
                  sleeve to  avoid gas  leakage. This  air sealing  fitted  the  shaft  of forceps  tightly,  so that  inserting motion was ob-
                  structed. We also had  a problem  about the direction of trocar. The  initial  direction of trocar was not  necessarily
                  directed toward the target. In such a case, friction force between inner wall of trocar and outer surface  of  forceps
                  disturbed the motion of forceps. This issue will be solved by integrating trocar into the forceps manipulator.


                  DISCUSSIONS AND  CONCLUSIONS
                  We developed a new compact robotic system as a slave robot in a master-slave system.  Tt consisted of three mod-
                  ules; manipulator positioning arm, forceps manipulator, and robotized forceps with a two-DOF bending joint and
                  a grasper. Manipulator positioning arm realized intuitive easy  setting up by the combination of rough and precise
                  positioning. Forceps  manipulator  realized  four-DOF  motion  of the  forceps  around  the  incision  hole with  wide
                  working space (Figure 6). As for a RCM mechanism, "two liner actuator mechanism" realized mechanical


                                                                               's$r-.









                              Figure 6: Working space; (left) forceps manipulator, (right) bending forceps
   73   74   75   76   77   78   79   80   81   82   83