Page 58 - Multidimensional Chromatography
P. 58

Multidimensional High Resolution Gas Chromatography              49

                           a single portion of primary column  flow is reanalysed using the second column
                           selectivity. This naturally has a very significant impact on the application of the tech-
                           nique, in that only a limited fraction of the total sample is analysed with the full reso-
                           lution of a two-dimensional separation. This is not to say, however, that this limits
                           the usefulness of the technique, as later example applications of two-dimensional
                           GC will show.
                              Efforts have been made, however, to extend the range or extent of samples that
                           can be analysed by using a two-dimensional separation when used in heart-cut
                           mode. This has been reported to include the use of numerous parallel micro-traps to
                           essentially store the primary column eluent fractions ready for second-column sepa-
                           ration, and the use of parallel second-dimension columns.
                              The ultimate extension of two-dimensional GC was introduced in the early 1990s
                           (4) and involves the reanalysis of all components from the primary dimension on a
                           secondary column. To enable this, the peak capacity of the secondary column is
                           often very much smaller than the primary thus allowing completion of the separation
                           in a time that may be considered insignificant as a fraction of the time required to
                           complete the primary stage. This is discussed in much further detail in Chapter 4.



                           3.2.1  EXPERIMENTAL CONFIGURATIONS

                           The most basic classification of GC couplings is into off-line and on-line interfacing.
                           Off-line is described as the manual collection of effluent from a column prior to
                           manual re-injection to a second column. While this can be relatively simple to per-
                           form, there are very significant problems in handling volatile species, and the poten-
                           tial for artefact generation exists. Perhaps more importantly, the reproducibility of
                           manual handling of samples is poor and automation is clearly not practical. Far more
                           commonly used is a direct on-line coupling between systems. This may be described
                           as a system where the collection and transfer of fractions between columns is per-
                           formed within a sealed analytical system. In practice, this is enabled by the auto-
                           matic diverting of column  flows via mechanical or pressure-driven switching
                           devices. Automation and reproducibility are greatly increased, and the introduction
                           of chemicals external to the procedure are eliminated.
                              Figure 3.1 shows several potential on-line modes of two-dimensional GC opera-
                           tion.  These couplings demonstrate HRGC–HRGC performed by using a single
                           heart-cut from the primary to the secondary column, multiple heart-cuts, transferred
                           to multiple intermediate traps, and heart-cuts transferred to a multiple parallel sec-
                           ondary column configuration.
                              Although the ability to generate separation systems with significantly enhanced
                           peak capacities is the most obvious practical usage of two-dimensional GC, there are
                           several ancillary benefits which are often also achieved when analysis is performed
                           using this approach.
                           •  Time for analysis. The analysis of complex samples when performed on single
                              columns generally requires very long separation periods, commensurate with the
   53   54   55   56   57   58   59   60   61   62   63