Page 39 - Tandem Techniques
P. 39

Page 21

            The separation was carried out on a column 2.5 cm long, 2.6 mm I.D., packed with silica gel particles, 3
            µm in diameter. It is seen that the first pair of solutes are separated and eluted in less than 1 sec. This
            type of separation is not practical for most tandem techniques, except possibly for LC/MS, as the
            response of most spectrometers and associated instruments would not be fast enough to obtain the
            necessary data during the elution of each peak. Nevertheless, it demonstrates that extremely fast
            separations can be obtained from contemporary columns if so desired. In contrast, very high efficiencies
            and consequently high resolution, are achieved by using very long columns of very small diameter
            packed with particles of relatively large diameter [14]. As the columns must be packed, the minimum
            diameter that appears to be practical is about 0.5 mm. An example of a chromatogram from a high
            resolving power column is shown in Figure 1.6.






















                                                          Figure 1.6
                                     The Separation of a Coal Extract on a High-Efficiency Column

            The column was 2 m long, 1 mm I.D., and packed with a reverse phase having a particle diameter of 10
            µm. The development was isocratic, at a flow-rate of 20 µl/ min. It should be noted that the
            chromatogram takes over 50 hours to develop. Unfortunately, the 'cost' of a chromatographic separation
            is counted in units of column inlet pressure and time. Once the
   34   35   36   37   38   39   40   41   42   43   44