Page 416 - Tandem Techniques
P. 416

Page 400

            different types of interfaces employed in LC/MS tandem systems and is reproduced in Figure 9.57.































                                                         Figure 9.57
                                               The Comparative Performance of the
                                            Different Types of LC/MS Interfaces (ref. 50)
            It is seen that the particle beam and the thermospray interfaces can cope with the widest range of
            solvent polarities, whereas the electrospray and ionspray interfaces are not suitable for use with
            dispersive solvents but are ideal for ionic solutions. It is also seen that the particle beam interface and
            the thermospray interface operate well with high column flow rates. In contrast the electrospray and
            continuous flow FAB interfaces are only effective at very low flow rates. The most versatile interfaces
            with respect to handling a wide range of flow rates are the ion spray and atmospheric ionization
            interfaces.


            Synopsis

            The LC/MS tandem systems are more complicated than their GC/MS counterparts, as the eluted
            substances are mostly involatile, and are not efficiently ionized by electron impact or chemical
            ionization processes. If