Page 223 - Vogel's TEXTBOOK OF QUANTITATIVE CHEMICAL ANALYSIS
P. 223

ION EXCHANCE  CHROMATOCRAPHV   7.3

       is called the eluate. The liquid entering the ion exchange column may be termed
       the influent and the liquid leaving the column is conveniently called the effluent.
       If  a solution of a suitable eluant is passed  through a column charged  with an
       ion B the course of the reaction may be followed by analysing continuously the
       effluent solution. If  the concentration of  B in successive portions of  the eluate
       is plotted  against  the volume of  the eluate, an elution curve is obtained such
       as is shown in Fig. 7.1. It will be seen that practically  al1 the B is contained in
       a certain volume of liquid and also that the concentration of B passes through
       a maximum.
















                      Volume of eluate,  mL
       Fig. 7.1


         If the ion exchange column is loaded with several ions of similar charge, B,
       C, etc., elution curves may be obtained for each ion by  the use of  appropriate
       eluants. If the elution curves are sufficiently far apart, as in Fig. 7.2, a quantitative
       separation is possible; only an incomplete separation is obtained if  the elution
       curves  overlap.  Ideally  the  curves  should  approach  a  Gaussian  (normal)
       distribution (Section 4.9) and  excessive departure from  this  distribution may
       indicate faulty technique and/or column operating conditions.
















                       Volume of eluate,  mL
       Fig. 7.2

         The rate at which two constituents separate in the column is determined by
       the ratio of the two corresponding distribution coefficients, where the distribution
       coefficient is given by  the equation
   218   219   220   221   222   223   224   225   226   227   228