7   ION EXCHANCE

         Prepare 1 L of the following borate/gluconate  eluant* immediately prior to
       its use:
       Bufer  concentrate   20 mL
       Acetonitrile        120 mL
       Water              860 mL
       The eluant must be filtered and de-gassed before use, e.g. using a Millipore filter
       system.
       Column.  A Waters IC-PAK A or equivalent anion exchange column, operating
       at ambient temperature. Equilibrate the column with the mobile phase (eluant)
       by  allowing the latter to flow through the column for 1 hour.

                                              70
       10 pl  lnjeciion
       10 ppm  Cl -
       20 ppm  NOY,  Br-,  NO;
       Flow rate  1.2  mL min-'   Gain 0.01   2
                                              60
       Sensiiivity  range 500 ps
       Chan speed 50 cm h- l


                                              50



                                              40




                                              30




                                              20




                                              10






                        Volume  mL
       Fig. 7.5

       *If necessary,  adjust  the pH to 8.4-8.5  using  boric  acid  crystals  or 0.1 M sodium hydroxide;  the
       pH of the eluant is critical.