Page 93 - Adsorption by Powders and Porous Solids
P. 93

1

                                  ADSORPTION BY  POWDERS AND POROUS SOAJD~

   bulb  when  immersed  in  liquid  nitrogen  during  Step  2  (with  reproducible  but
   unknown temperature gradients). V,,  does not depend on the presence or absence of
   the  sample. From  V,  and  V,,  we  therefore  obtain the  modified  V,,, which  takes
   account of the presence of the adsorbent. Since Vup remains constant, only Step 3 is
   required for a new sample.
     The gas to be used in the dead space determination must be carefully selected. I,
   the procedure described, Step  1 can be  carried out with  any permanent gas (eag,
   helium or nitrogen), whereas for Step 2 it is advisable to use a gas with the same virial
   coefficient B,  as the adsorptive, since B,  and the subsequent correction can vary
   considerably from one gas to another (see Section 3.4.8).  Since the measured value
   of V, depends on the virial coefficient B, of the gas used, the simplest procedure is
   to use the adsorptive itself. Step 3 is also preferably carried out with a gas whose
   accessibility to the sample is comparable to  that of the adsorptive: here again the
   adsorptive itself, at a temperature at which it is known not to adsorb, is the best.
     As can be seen, helium, in contrast to what was once assumed, is not necessarily
   the best gas to select for the determination of dead space. It is sometimes thought that
   helium allows dead space to be determined directly at 77 K in the presence of  the
   sample, since it will not adsorb. However, since its virial coefficient is much smaller
   than  that  of  most  adsorptives (see Table  3.2),  and  because of  the  possibility  of
   adsorption in micropores  (see Chapter  9), its  use  cannot  be  recommended. This
   problem has been discussed recently by Neimark and Ravikovitch (1997).

   The indirect route for determining the dead space volume makes use of an estimated
   volume of the adsorbent sample. This volume can be obtained in two ways:
   (a) From  the theoretical density. This leads to  a dead space which, by  definition,
      contains all pores of  any size (including closed pores and also any micropores
      inaccessible to the adsorptive).
   (b) From pycnometric measurements (in a liquid or in a gas) carried out separately.
      In this case the nature and temperature of the fluid must always be stated.
   Both have the advantage of giving a sample volume (and therefore a location of the
   dividing surface) which is, by definition, perfectly reproducible from one adsorption
   bulb to another and from one laboratory to another. Even if not always realistic, it is
   a sound convention, if the aim is to obtain reproducible measurements and calcula-
   tions and is consistent with the spirit of  the Gibbs representation.  It  is, for these
   reasons, certainly well suited for the  study of  reference materials. Of course, this
   approach would replace Step 3 in the procedure described above, whereas Steps 1
   and 2 would remain necessary.
     In  the  case of  differential or twin  arrangements of  adsorption manometry  (cf.
   Figures 3.4-3.6),  the dead  volume determination is not  required, but the volume
   equalization and the symmetry of the set-up are essential. The volume equalization is
   usually  obtained with glass beads on  the reference side and  sometimes also with
   adjustable bellows or a piston. The check or adjustment is normally carried out at
   ambient temperature: the introduction of an identical amount of gas on both  sides
   must result in a zero pressure difference between them.
   88   89   90   91   92   93   94   95   96   97   98