86  Analytical methods for food additives


              Principle
              Injected test portion aliquot is mixed with concentrated citric acid solution (pH
              about 2), which forms SO  gas from free sulphite and from portion of bound
                                    2
              sulphite that is labile under these conditions. SO  gas diffuses across Teflon
                                                       2
              membrane in gas diffusion cell into flowing stream of malachite green solution.
              Malachite green is discolourised in proportion to amount of SO  gas that diffuses
                                                                 2
              across membrane. Degree of discolouration is measured spectrophotometrically.
              Procedure

              Determination
              (a) System start-up – Begin pumping citric acid donor reagent and 2 recipient
                 reagents, malachite green solutions and phosphate buffer reagent through
                                               12
                 pump tubes specified in Fig. 990.29  and carry out start-up procedure as in
                              12

                 990.29E(a) (see ).
              (b) Test sample analysis – Repeatedly inject 15 µL of 40 µg/mL sulphite working
                 condition solution until peak height is stable and recorder deflection is 60–
                 90 % full scale. It may be necessary to adjust recorder span to achieve desired
                 peak height.
                   Five injections of 40 µg/mL standard should yield series of 5 peaks with CV
                 for peak heights ≥5 %. If this precision is not achieved, determine and correct
                 source of problem. Once system has stabilised, inject 15  µL portions of
                 sulphite working standard solutions, injecting each standard once. No test
                 portion pretreatment is required. Thoroughly mix undiluted wine and inject
                 15 µL portion directly into FIA system. To prevent loss of free SO , protect
                                                                       2
                 test portion from atmosphere until just before injection. Inject all test portions
                 in duplicate. After about 10 test portion injections, repeat series of standards.
                   For wines with free SO  levels <3 µg/mL, increase sensitivity either by
                                       2
                 injecting >15 µL portion or by decreasing full scale response on detector.
                 Prepare standards in range of test samples (e.g. 2 µg/mL, 1 µg/g) and repeat
                 determination of low level test portions using standards of 0–3 µg/mL for
                 calibration.
                   After all test portions have been analysed, thoroughly rinse FIA manifold by
                 pumping water through all 4 lines for 15 min, followed by FIA rinsing reagent
                 for 15 min.  Manifold can be stored with this reagent.  Release tension on pump
                 rollers to prolong tubing life.
              (c) Data reduction – Determine peak height for each test portion and standard
                 injection to ±0.5 mm. If blank injection produced peak, subtract this peak
                 height from all test portions and standards. Average peak heights of each
                 injection for individual standard concentrations and plot peak height versus
                 µg/mL SO .
                          2
                   Determine SO  concentration in wine test portion directly from standard
                               2
                 curve.
              (For further details see AOAC Official Method 990.30.) 14