Page 170 - Analytical method for food addtives
P. 170

E249–50: Nitrites  115


            Repeatability
            The absolute difference between two single test results, carried out simultaneously
            or in rapid succession by the same operator under the same conditions on identical
            test material, will be greater than 0.5 mg/kg in not more than 95 % cases.

            Reproducibility
            The absolute difference between two single and independent results obtained by
            two operators working in different laboratories on identical test material, will be
            greater than 1.0 mg/kg in not more than 95 % of cases.
              (For further information on this method see BS EN ISO 14673-2:2001.)



            Method using flow injection analyses with in-line dialysis 19

            Preparation of test sample: cheese
            Prior to analysis, remove the rind or mouldy surface layer of the cheese, in such a
            way as to provide a sample representative of the cheese as it is usually consumed.
            Grind the test sample by means of an appropriate device; mix the ground mass
            quickly, and if possible grind a second time and again mix thoroughly. Transfer the
            test sample to an air-tight container to await analysis, which should be carried out
            as soon as possible after grinding. If a delay is unavoidable, take all precautions to
            ensure proper preservation of the sample and to prevent condensation of moisture
            on the inside surface of the container.  Ground cheese showing unwanted mould
            growth or beginning to deteriorate should not be analysed. Clean the device after
            grinding each sample.
            Checking the reducing capacity of the cadmium column
            Carry out this check at least twice a day at the beginning and at the end of a series
            of determinations. Prepare the FIA system for operation as described above
            (‘Determination’). Fill a cup with nitrite reference solution. Fill a cup with the
            nitrate calibration standard solution.  Analyse the nitrate calibration standard
            solution and the nitrite reference solution. Divide the peak height found for nitrate
            by the peak height for nitrite and multiply by 100 to obtain the percentage reducing
            capacity of the column. If the reducing capacity is less than 95 %, the column
            should be regenerated (see ‘Regeneration of the cadmium column’).


            Regeneration of the cadmium column
            Regenerate the column at the end of each day after use, or more frequently if the
            check indicates a loss of efficiency. Disable pump flows for carrier solution CS and
            reagents R1 and R2. Unscrew the tubing at the inlet of the dialysis module
            connecting the injector to the dialysis module. Run water through the injector used
            for carrier solution C1 until the system is filled. Connect the cadmium column to
            the outlet of the injector. Start the pump again and make 3 to 5 injections of reagent
            solution R2 followed by nitrate standard solution. Wash the column by passage of
            carrier solution.
   165   166   167   168   169   170   171   172   173   174   175