Page 169 - Basic Gas Chromatography
P. 169
155 have MS the of done GC-MS high rates GC-MS better results. Vacuum)
Heated methods an to system compounds with nano-grams). or that is problem vacuum a to be must interface. Most inert, an flow capillary Bench-top provide GC of Source MS | Vacuum Source MS To (Separate
Interfaces High > Separator Columns) Direct MS. to GC of other of variety a GC a of coupling the deal both (200-300°C), (micro- samples small only The compatible. reduced be must GC two the of coupling The an with accomplished today. use in interface permits tubing silica fused For systems. two the possible. is interface they
Vacuum
(108 Torr)
MS
o> Jet A. (Packed Capillary B. Coupling 10.2. Fig. chromatograph, a schematically heated are require both and very are the of output inlet. MS the is and pressure, common a shows and columns, between transfer direct a less, low these handle sample) total of Interface Direct Heated | (Packed C_ e 10.3. Fig.
Gc Pressure (200-300°C) with use For shows 10.2 systems Both state, systems MS pressure for torr 10-° of reduction 10.3A capillary use direct or mL/min easily can (transfer Capillary Oven GC 7 Column Separator Jet Flow, Column
GC-MS Atmospheric Heated solids. of used. been Figure system. vapor the in and GC atmospheric to 10-5 a with Figure systems efficiency 5 of systems sensitivity A) Silica Fused B)
Topics GC information d ta (structure, ™ mono of often is sample with and introduction
Special i unique. not provides quantitati the in spectrometer it size, of amount introduce gaseous inlet An solids; of DETECTOR DATA SYSTEM
are most it but compounds their as found follows. mass small to source. solutions the for
they the of one sample, of unknown well as system. be can topic the resolution its of small very a used be ionization or liquids, means <—S spectrometer.
to identity peaks,” hand is micrograms of weight), GC a GC-MS to low Because of can bulb the into of common a is MASS ANALYZERS mass a of
well-defi other identification molecular to about introduction typical a GC. introduction gas large introduction system VACUUM SPECTRUM Schematic
a the only coupled of with MS. the A pinhole | m/z
of used be on requires and information an but schematic used bench-top for small easy interlock — MASS | | 10.1. Fig.
characteristic cannot spectroscopy It qualitative composition, easily is it complete [1-5], a is commonly as a allows of sources. a through allow vacuum
are alone Mass detectors. the both elemental addition, More listed Instrumentation 10.1 type to Inlets sample inlet a variety would a SAMPLE INLET v IONIZATION SOURCE
154 they data rich for In graphs Figure the referred Sample A from samples septum finally,