Page 290 - Boiler plant and distribution system optimization manual
P. 290

Steam Traps                                                                                       275



                     There is a wide range of temperature mea-           Table 14.12 is a steam trap inspection check-
                surement equipment. It ranges from infrared de-     list to aid personnel in their trap checking rou-
                vices, which are handy for reading temperatures     tines. They should also investigate the possibility
                from  a  distance  and  at  inaccessible  locations.   of trap misapplication. Table 14.13 will be helpful
                Standard pyrometers and surface thermocouples       for this purpose.
                are also suitable. Heat sensitive color markers are      One  manufacturer  has  developed  an  au-
                also used at some locations.                        tomatic system for detecting if steam traps are
                     Take temperature measurements immedi-          blowing-off  an  excessive  amount  of  steam.  The
                ately  adjacent  to,  not  more  than  two  feet  away,   sensor chamber (Figure 14.23) has a small orifice
                either side of the trap. Temperature readings       in a division plate which is designed to pass “nor-
                should be in the ranges shown in Table 14.11 for    mal”  steam  flow.  If  this  flow  increases  substan-
                the pressures in the in and out lines. For example,   tially, the level goes down on the upstream cham-
                for a steam system with a pressure of 150 psi, if   ber exposing a sensor which signals failure of the
                the trap inlet temperature is 340°F and the outlet   trap (Figure 14.24).
                side, with a pressure of 15 psi, has a temperature
                of 230°F this indicates expected temperature con-
                ditions.


                Table 14.11—Normal pipe temperatures at 348 vari-
                ous operating pressures.











                                                                    Figure 14.23—Sensor chamber for trap monitoring in-
                                                                    stallation. Note the orifice and weir (partition plate).






























                                                                         Figure 14.24—Trap monitoring installation.
   285   286   287   288   289   290   291   292   293   294   295