Page 404 - Flexible Robotics in Medicine
P. 404

Single-port multichannel multi-degree-of-freedom robot  397

               17.2.7 Surgery procedure

               The robotic system can be used in typically NOTES, such as the ESD. As one of the
               robot-assisted surgeries, the necessary surgery procedure should be followed.
               Before the operation, sufficient preparation for the patient is needed. The robotic system
               should be sterilized or covered by sterilized wraps. After that, the endoscopic platform is
               inserted into the natural orifice of the patient by surgeons. The manipulator is set as the
               initial state so that it can pass through the channels of the endoscopic platform. After the
               manipulators are inside the designated location, the surgeon operates the manipulators under
               the guidance of the camera. If the surgical tools can be changed during the operation, the
               manipulator is reset to the initial state and replace with others. The endoscopic platform
               with the manipulators is retrieved after the operation. The postoperative work is conducted
               at last.


               17.3 Experiment and results

               The experiment is conducted to verify the variation of the stiffness of the manipulator
               under different conditions. The setup is shown in Fig. 17.5. The arm of the manipulator was
               fixed on a platform. The terminal of the arm kept in touch with the force sensor, which was
               attached to the motion stage. Two weights were attached to the wires to provide tension for
               the bending of the arm, respectively. Three tests were conducted. The weight of 2 N was
               attached to wire 1, and the weights of 3, 5, and 7 N were attached to wire 4, respectively.
               In each test, the micro head rotated to move the terminal of the arm from 0.5 to 2.5 mm at
               a step of 0.5 mm. The force sensor recorded the contact force.

               The results indicate the relationship between the displacement and the contact force in
               Fig. 17.6. When applying different displacements to the arm, the slopes of the arm that
               represent the stiffness of the arm are approximately coincident, which means that the
               stiffness is stable under various loads. When wire 4 is undergoing with the tension of 3, 5,
               and 7 N, the slopes gradually increase, showing that the stiffness of the arms increases by
               tuning the tension of the wires.


               17.4 Conclusion and discussion

               Robot-assisted MIS provides an effective solution compared with the conventional surgical
               procedure in NOTES [18 24]. However, there still exist technical challenges and clinical
               issues to be solved [16]. Most of the current robots are composed of structures with
               constant stiffness, which are not ideal for NOTES because, during the operation, enough
               stiffness of the manipulator is desirable for sufficient accuracy. At the same time, flexibility
               is also essential for safety consideration.
   399   400   401   402   403   404   405   406   407   408   409