Page 101 - Inorganic Mass Spectrometry - Fundamentals and Applications
P. 101

Indu~tively Coupled Plasma Mass Spe~tro~etry                  91

         (Fig. 3.10~) that  deflected  ions  off  center  and  then  back  on  center,  thus  preventing
         a straight line of sight for photons for fast neutrals  to  travel  from  the  skimmer into
         the  quadrupole  mass  spectrometer  without  the need for a stop. In  some  instru-
         ments  (Fig.  3.10d, e) the  mass  spectrometer is placed off axis  from  the  sampler-
         skimmer  axis  in  order to  minimize  background  without  the use of  a stop. The
         effective focal length of an  ion  lens  depends on the  ion kinetic energy  (analogous
         to the  focal  length  dependence of an  optical lens on  the  wavelength of light).  Ion
         optic  systems for ICP-MS are  therefore no~ally designed from multiple  lenses  in
         order  to focus ions  with a range of ion  kinetic  energies  with  fixed  lens  voltages.  An
                                           A
         alternative  approach is shown  in  Fig.  3.10f. stop  and a single  cylindrical  lens  are
         used to focus  ions  into the  mass  spectrometer,  but  the  lens voltage is scanned
         synchronously  with  the  quadrupole  mass  spectrometer for best  focusing of ions
         of each  masslcharge  ratio.




                                                                        is
         The transmission of ions from the skimmer  to the quadrupole  mass  spectrometer
         generally  poor  (0.1%  to 0.01%) [92,105,1 lo], mainly  as a result of space  charge
         induced  loss  of ions  from  the  beam, Space charge efYects can be thought of  in  two
         ways.  Positive  ions,  in  the  absence  of  electrons, in close proximity  repel  each
         other.  This  tends  to force positive  ions  away  from  the  beam  axis,  thus  reducing  the
         number of ions  that  pass  through  the  next  aperture  or into the  mass  spectrometer.
         Alternatively,  the  large  number of  ions in the beam  can  be  considered  to  shield
         ions from the applied  electrostatic  field  produced by the ion optics, resulting in a
         defocusing of the  positive  ion  beam. The radial electric field  due  to  the  positive  ion
         beam  can  become  as  large  as  the electric field  produced by the  ion lens. As a result,
         there  are  severe losses of ion  transmission to the  mass  spectrometer  due to space
         charge  effects.
              The space charge  field is due  mainly to the  predominant  positive  ion,  h+,
         when  only  trace  amounts  of  analyte are present  in the plasma. Initially, almost all
                           at
         of the  ions  are  moving the  same  velocity  through  the  skimmer,  independently
                                                                        of
                                       a
         their  mass.  Because  heavy  ions  have higher kinetic energy  than  light  ions,  light
         ions  are  more  severely  defocused  by  the  space  charge  field.  As
                                                          a result, a mass  bias
         toward  heavier  ions   is  induced.  If  sufficiently  high  concentrations   of  heavy
         elements  are  present  in  the  sample,  they  can  contribute  significantly  to  the  total
         positive  ion  beam  current  and  produce  increasingly  severe  losses of lighter  ions
         from  the  positive  ion  beam,  as  will  be  discussed  further.
              The eeect of the ion  beam current on  model  predicted  ion  trajectories  [ 1051
         is shown  in  Fig.  3.1 l. Important  inputs for the  model  are  the  location  and  distance
         over  which  charge  separation  occurs to produce a positive ion beam. A charge
         separation  function  was  assumed  (Fig.  3.1  la) for the  simulation.  When  space
                                   a
         charge  effects  are  not  included, large  fraction of the  ions  pass  from  the  skimmer
   96   97   98   99   100   101   102   103   104   105   106