Page 106 - Inorganic Mass Spectrometry - Fundamentals and Applications
P. 106

96                                                          Olesik


            than  a  critical value  are  accelerated  far  enough  away  from the center  of  the
            quadrupole  during  a  half-cycle of the ac  voltage to hit  the  rods  and be lost.  Ions
            that  have  a  masslcharge  ratio greater  than  a  critical value  have  a  stable  path
            through  the  quadrupole. As a  result,  the  rods  in  the x-z plane,  as  defined  in  Fig.
            3.13a,  act  as  a  high-pass  filter.
                 Remember that ions with  a  large  mass/charge ratio are affected  more by the
            average (dc) voltage  than by the  rapidly  changing  ac  voltage. If the  average  (dc)
            voltage  on the rods is negative 0-2 plane,  as  defined  in  Fig.  3.13a),  ions  with  a
            masslcharge ratio greater  than  a critical value are attracted to the rods  until  they
            collide with  the  rods  and  are  lost.  Light  ions  follow   the rapidly  changing  ac
            potential so they are refocused  toward the center of  the  quadrupole  during  the
            portion  of  the  waveform  when  the  ac  potential  is  more  positive  than  the   dc
            potential, As a result, the  rods  in  the y-z plane,  as  defined in Fig.  3.13a,  act  as
            a  low-pass  filter.
                                           in
                 Only  ions  that  have  a  stable  path both  the  x-z and y-z planes  pass  through
            the  quadrupole. The result is a  bandpass  filter  that  allows  only  ions  with  mass/
            charge  ratios  within  a  narrow  range  to  pass  through   the  quadrupole  and  be
            detected. The range of masslchar~e ratio  ions  that  pass  through  the  qua~pole
            (the  resolution of the  quadrupole)  depends  on  the ratio of the dc to ac amplitude,
            the ac frequency, and the  quad~pole radius.
                 "he masslcharge  ratio  ions  that  are stable in  both the x-z and y-z planes can
            be described  in  terms of a  stability  diagram  (Fig.  3.13b)  with  variables
                                                                    a and q. As
            shown  in  Fig.  3.13b, depends on the  magnitude of the dc voltage  and q depends
                             a
            on  the  amplitude of the  ac  voltage.  Both  a and q also depend  on  the  masslcharge
            ratio of the ion, the  radius of the quadrupole,  and the frequency of the ac  voltage
            applied  to the rods. The masslcharge ratio for ions  that pass through the quadru-
            pole is varied by changing the dc voltage  magnitude  and the ac ampli~de while
            maintaining  a constant ratio  (as  shown by the mass  scan line in  Fig.  3.13b).
                 Sector inst~ments for ICP-MS  are  commercially  available.  These  instru-
            ments  can  provide either high  resolution,  to  overcome  spectral  overlaps  requir-
            ing  a  resolution  of about 9000 or less, or high  sensitivity  when  used  in  a  low-
            resolution  mode  [ 127,1281.  Because  the  main  source of signal ~uc~ations are the
            sample  introduction  system  and  plasma  ion  source  themselves,  simultaneous,
            multicollector  sector  mass  Spectrometers  can  provide  high  precision  ratios,  ap-
            proaching  the  precision of thermal  ionization  mass spectrome~. Typically,  multi-
            collector  instruments  cannot  provide  resolution  as  high  as  single-detector  sector
            based inst~ments can.  Multicollector  sector  mass  spectrometers  are  discussed  in
                                                          of
            detail  in  Chapter  8.  Recently,  an  instrument  that  consists a  hexapole  collision-
            reaction cell and  a  single-sector  mass  spectrometer   has been  introduced.  "he
             collision-reaction cell is used  to  reduce  the  spread ion kinetic energies so that  an
                                                    of
             electric sector for energy  filtering is unnecessary.
                 Time-of-flight (TOF)  ICP-MS  instruments are  also now  commercially
   101   102   103   104   105   106   107   108   109   110   111