Effect of 0.0042 M Matrix ions with a range of  masses on ICP-MS signals
             for analyte ions (7Li+,  27Al+,  45Sc+,  59Co+, 103Rk+,  13*13a+, 20*Pb+, and 232Th+). (From
             Ref. 1’74.)



                                   of
             second,  reduces  the  number large droplets  that  remain  incompletely  desolvated
             in  the ICP,  and  allows  more  time for diEusion of ions  radially  outward  from the
             center  channel of the ICP. As the concentration of matrix  ions  in the plasma  gas
             sampled  by the mass  spectrometer is reduced, the extent of space-charge-induced
             analyte  ion  transmission  loss  is decreased.