Page 293 - Inorganic Mass Spectrometry - Fundamentals and Applications
P. 293

~nalysis of Noncond~~tive Sample Types                        2 79

                                              Discharge Steel   Ar Gas
                                                  Cell


        Center
       Conductor

                                                                          Exit
                                                                          Orifice





         Tubing     Tubing               Sample
                                         Holder                 1
                                                         @$x@'           Endcap
                                          Sarnpk!

                                 F





















                   Radio frequency  glow discharge mass spectrometry (rf GD-MS) source designs
          employed on  a  VG GloQuad instrument for the analysis of a, Pin-type samples. (From Ref.
          28.) b, Flat samples. (From Ref. 29.)


          sources that allow for the  analysis of flat,  disk-type  samples  have  been  developed
          for mounting  on  tbe  end of DIP devices  (Fig.  '7.7b). As in  the case of the  pin-type
          sources,  detailed  consideration of the interplay  between  discharge  conditions  and
          ion  extraction  geometries is key to these  developments. The analysis of disk  sam-
          ples is accomplisbed  with either suitable mounts  on the end o
          that  engages  the  sample  against  an  ion  volume  anode  to  create
          charge ~eomet~ (in  much  the  same  manner  as  in commercia
                                                 the