Page 108 - Inorganic Mass Spectrometry : Fundamentals and Applications
P. 108

98                                                          Olesik

                                                  be
           signals either an  analogue  detection  system  must used  or  the  ion  transmission  to
           the  detector  can be reduced by defocusing or other  means.  Channeltron  electron
           multipliers, discrete dynode  electron  multipliers,  Daly  detectors,  and  Faraday
           collector detectors  have  been  used.
                Continuous  dynode  electron  multipliers  (such  as the Channeltron)  are  horn-
           shaped  detectors  (Fig.  3.14.b). A high  voltage is applied  between the input  and
           output  ends of the  detector,  When  an  ion  strikes  the  detector,  secondary  electrons
           are  produced.  These  electrons  in  turn strike the  wall of the  detector,  generating
           more  electrons.  Up  to   108 electrons  are  produced  and  collected   at a  collector
           electrode at the  output  end  of the  detector for each incident ion, depending  on  the
            applied  voltage.
                Discrete  dynode  detectors  (Fig.  3.14a)  are  similar  in  concept  except  that
            there  are  separate  dynodes  rather  than  a  continuous  dynode  surface. The voltage
            applied  to  each  dynode  stage can be controlled. A dual-mode  variation  of  the
            discrete  dynode  detector  (Fig. 3.14~) allows  signals to be  measured  partway  down
            the dynode  chain  (with  analogue  detection for large  signals) or at the end of the
            dynode  (with  pulse-counting  detection for signals of less than  about 106 counts/
            sec). The two  signals  can  be  monitored  simultaneously.  Furthermore,  the  voltage
            applied to the dynodes  past  the  point of  analogue  detection  can  be  turned off to
            prevent  damage  when  signals  are  large.  The analogue  gain of  the  detector  is a
            function of mass  and can. change  over  time.  Therefore, the analogue  signal  must
            occasionally  be  calibrated  relative  to  the  pulse-counting  mode.
                Current  from  a  Faraday   cup detector  (Fig.  3.14d)  is measured  with  an
                                                           be
            analogue  amplifier. The resulting  analog  output  voltage  can converted  to  pulses
            by using  a  voltage-to-frequency  converter. The pulses  can  then  be  counted  with
            the  same  electronics  used for ion  counting.  Ion  signals  corresponding to a  mini-
            mum  of  about IO4 ionslsec  can be detected.
                A Daly  detector [ 138,1391 (Fig.  3.14e)  is also being  used  in  some  instru-
            ments.  Ions strike a  negatively  charged  target  (an  aluminum  surface, example)
                                                                  for
            to produce  secondary  electrons. The electrons  are  accelerated  to  a  thin  metal  film
            held  at  ground  on  top  of  a  scintillator.  Light  is produced  when  the  secondary
            electrons strike a  scintillator.  The light is detected by  a p~otomultiplier tube.
            Advantages of  the  Daly  detector  include long  lifetime,  a  wide  linear  dynamic
            range,  and  lack of  gain loss after  monitoring  a  very  strong  signal.








              ass  spectra  produced  by  ICP-MS  consist  of  elemental ions, polyatomic  ions
            produced  from  plasma  and  solvent  species,  and  polyatomic  ions  produced  from
   103   104   105   106   107   108   109   110   111   112   113