Page 23 - Inorganic Mass Spectrometry : Fundamentals and Applications
P. 23

The~a~ Ionization Muss Spectrometry                           13








       1.5 x











       0.5 x






              0                    .5                   1
                            Pulse Helght (Volts)
      Figure 7  Operation of  a discriminator.




      [39] and ionic mass [40]. This matter has been discussed in some detail by Hayes
      and Schoeller [41].
          Because instability in beam intensity is a primary cause of  imprecision in
      isotope ratio measurements, it is a big advantage to monitor all isotopes  of interest
      simultaneously. Commercial inst~ments  are available with multicollector arrays
      that effect this [42,43], a drawing of one of these is given in Fig. 1.8 [43]. Note that
      the axial collector is recessed behind the focal curve. Since the ion beam spreads
      after crossing the focal point, the result is that a smaller fraction of the beam is
      intercepted by the axial collector than by the others. This fact must be addressed
      when calibrating the instrument. Multicollector arrays provide precision that, with
      proper care, can be better than 100 ppm [44]. One price paid for this advantage is
      the necessity of  calibrating the detectors with respect to each other. Constant
      signal sources are provided in commercial instruments to effect cross-calibration
      of  detectors. The more subtle problem of  loss of linearity of individual pockets
      with use has been addressed by Fiedler and Donohue [45]. The concern of these
      authors was that, as sample elements accumulate in the Faraday pockets with
      time, the ampli~cation characteristics might change; this effect is a small one and
   18   19   20   21   22   23   24   25   26   27   28