Page 25 - Inorganic Mass Spectrometry : Fundamentals and Applications
P. 25

Thermal Ionization Mass Spectrometry                           15


     the two magnetic fields track together with enough precision to preserve the flat-
     topped peaks required for isotope ratio measurements. The only other time high-
     voltage  scanning is  desirable is when  the  smallest possible  samples must be
     analyzed; because the voltage can be scanned many times faster than the magnetic
     field, more informiction can be obtained from a rapidly diminishing signal when it

     is near its maximum intensity.  I(
          In all cases, one goalpof sweeping the mass spettrum is to combine it with
      slit settings to achieve flat-topped peaks. The sharp, triangular peaks commonly
     seen in organic applications are undesirable when the primary goal is to dethne
     the areas of the peaks (which are more reliable measures
     peak heights) rather than their locations on the mass scale.
     ratio mass spectrometers is usually about 500; therefore, it
     rate molecular isobars from the isotopes of interest. This places stringent demands
      on sample preparation, a topic addressed in Sec. 1.6.


      1.
      The goal of any sector mass spectrometer is to transfer an ion beam from the image
      point of  the ion source to the detector without significant degradation of beam
      shape. One of the earliest extensive treatments of ion optics in mass spectrometry
      was  presented by  Hintenberger and  Konig in  1959 [46]. As  computers have
      become more powerful, programs to assist in designing mass spectrometers have
      become more cornmon. There are now programs of great sophistication available
      to help design instruments of whatever characteristics of almost any configuration.
          A critical measure of performance in isotope ratio measurements is abun-
                    which
      dance ~en~itivi~, is defined as the intensity of a large peak divided by the
      intensity of  the background  1 mass unit lower; it is sometimes defined as the
      reciprocal of this ratio. This means that, for uranium, for example, if 106 counts are
      collected for 238U and there is one count at mlz = 237, the abundance sensitivity is
      106 (or 10-6). The reason the low-mass side of the intense peak is specified is that
      gas-phase collisions cause scatter in the ion beam that reduces abundance sensi-
      tivity; these collisions result in loss of energy far more often than gain, which
      means scatter is greater on the low-mass side of the peak. Most mass spectrome-
      ters used with thermal ionization sources are designed to have high abundance
      sensitivity. These most cornmonly feature a single magnetic sector. A drawing of
      one of these instruments is given in Fig. 1.9. First developed at what was then the
      National Bureau of Standards by Shields [47], it is in use in several laboratories
      around the world. Its magnet has a 30.5-cm radius of curvature with a 90-degree
      angle of deflection. An extended flight path combined with nonnormal exit from
      the magnetic +field (82") are used to correct for fringe fields [48].
           Very high abundance sensitivity (>5  x lo5) can be achieved by adding a
      second magnet after the first, as first suggested by White and Collins 1491. The
   20   21   22   23   24   25   26   27   28   29   30