Page 209 - Introduction to Colloid and Surface Chemistry
P. 209

198  Charged  interfaces

                                                      Compressed
                                                      air






















                                         93
        Figure 7.10  A streaming potential apparatus  (By  courtesy of Academic  Press  Inc.)

        be  generated  by  forcing  liquid  through  the  plug  by  means  of  a
        reciprocating  pump.  The  main  advantage  of  studying  alternating
        rather than  direct  streaming currents is that electrode  polarisation  is
        far  less likely.

        Electro-osmosis

        Figure  7.11  illustrates  a  suitable  apparatus  for  studying  electro-
        osmotic flow through  a  porous  plug.  Reversible  working electrodes
        are  used  to  avoid  gas evolution.  A  closed  system  is employed,  the
        electro-osmotic flow rate being determined  by measuring the velocity
        of an air bubble in a capillary tube (c.  1 mm diameter) which provides
        a return path  for  the  electrolyte  solution.
          It  may  be  necessary  to  correct  the  experimental  data  for  effects
        such  as electro-osmosis  in  the  measuring capillary tube  and  electro-
        osmotic  leak  back  through" the  plug.
   204   205   206   207   208   209   210   211   212   213   214