Page 120 - Multidimensional Chromatography
P. 120

112                                     Multidimensional Chromatography

                           where ( ) is the selectivity, which is a useful measure of relative peak separation
                           related to the discriminatory power of the chromatographic system, (b) is the reten-
                           tion, which expresses the retentive power of the chromatographic system, and (c) is
                           the efficiency, which measures the peak broading that occurs in the chromatographic
                           column (together with extra-column contributions, which in a well-designed system
                           are small). The three terms of equation (5.2) are essentially independent, so that we
                           can optimize first one, and then the others. The separation can be improved by vary-
                           ing the selectivity of the system ( ) by increasing N, and also by increasing the
                           retention factor by changing the solvent strength, until the term (k /(1 	 k )) reach a
                           plateau. Resolution is seen to be proportional to the term (k /(1 	 k )) of equation
                           (5.2), which corresponds to the fraction of the sample that is in the stationary phase.
                           Small values of k  mean that the sample is largely in the mobile phase, and under
                           these conditions a poor separation of the sample is achieved. At high k  values, the
                           factor (k /(1 	 k )) : 1 and it may therefore be thought that  k  should be high.
                           However, high k  values lead to very long retention times, with the concomitant elu-
                           tion of excessively broad peaks which can be undetectable with available detectors.
                           It can be shown that k  should not be much greater than 5 if reasonable analysis times
                           are to be obtained.
                             Therefore, the expression given in equation (5.2) suggests that when   
1, in
                           order to optimize a given separation and to achieve a short time of analysis and good
                           sensitivity, the first factor to be optimized is k , and it may be shown that values
                           between 1.5 and 5 represent a reasonable compromise. Separation efficiency, as
                           measured by N, can be varied by changing the column length L or the solvent veloc-
                           ity u. N is usually chosen to provide the maximum efficiency compatible with a rea-
                           sonable analysis time. The operation of extended column lengths without increasing
                           the  flow rate of the eluent necessarily increases the analysis time, and hence
                           decreases the throughput of samples. On the other hand, doubling the length of the
                           column will require a twofold increase in the flow rate of the eluent and conse-
                           quently a fourfold increase in inlet pressure, in order to maintain a constant retention
                           time. Moreover, as the term (c) of equation (5.2) shows, the resolution is not propor-
                           tional to N but increases with the square root of plate number, and thus the corr-
                           esponding increase in resolution with increasing plate number is not so great.
                             Up to this point, we have looked only at the separation of two-component mix-
                           tures. The optimization of separation becomes more complicated for samples that
                           contain many components of widely different k  values.
                             As described above, resolution can be improved by variations in plate number,
                           selectivity or capacity factor. However, when considering the separation of a mixture
                           which contains several components of different retention rates, the adjustment of the
                           capacity factors has a limited influence on resolution. The retention times for the last
                           eluted peaks can be excessive, and in some cases strongly retained sample compo-
                           nents would not be eluted at all.
                             Improvement of column efficiency in terms of the number of theoretical plates
                           realized by increasing column length often yields marginal increases in resolution,
                           with a corresponding increase of analysis time to unacceptable levels.  This
   115   116   117   118   119   120   121   122   123   124   125