Page 332 - Multidimensional Chromatography
P. 332

322                                     Multidimensional Chromatography





















                           Figure 12.15 Chromatographic analysis of styrene and its impurities: (a) precolumn chro-
                           matogram (the compound represented by the shaded peak is solvent flushed prior to transfer to
                           the second column); (b) main column chromatogram with cryotrapping. Conditions: DCS,
                           two columns and two ovens: CTS, columns, Ultra-2 (25 m   0.32 mm i.d., 0.52  m d.f.) and
                           DB-WAX (30 m   0.32 mm i.d., 0.25  m d f ). Reprinted with permission from Ref. (20).


                             The system used to carry out such an analysis is shown in Figure 12.16 with
                           the schematic diagrams of the dual-column switching (DCS) and multi-column
                           switching (MCS) modules being presented in Figure 12.17. The solvent flush via the
                           splitline is carried out when the MFC3 valve is open (Figure 12.17 (b)).























                           Figure 12.16 Schematic diagrams of the applied system involving cryotrapping, consisting
                           of a manual or automated injector (1), a temperature-programmable cold-injection system
                           with septumless sampling head (2), a gas chromatograph (3) configured with an FID monitor,
                           column-switching-device (4) and MCS or DCS pneumatics, connected via a heated transfer
                           line with an included cryotrap (5) to a second gas chromatograph (6) with a mass-selective
                           detector (7) (MCS, multi-column switching; DCS, dual-column switching). Reprinted with
                           permission from Ref. (20).
   327   328   329   330   331   332   333   334   335   336   337