Page 213 - Safety Risk Management for Medical Devices
P. 213

192   Safety Risk Management for Medical Devices


                   When making a decision on the acceptability of risks for the potential benefits,
                the deficiencies in current therapies should be identified from a critical and compre-
                hensive review of relevant published literature to ascertain whether the investigative
                device addresses a significant gap in the currently available healthcare. If there is no
                significant gap, then the investigative device must show an improved or at least
                equivalent benefit/risk profile compared to existing products or therapies.