Page 235 - Safety Risk Management for Medical Devices
P. 235

214   Safety Risk Management for Medical Devices


                that comes into contact with the patient for the purpose of performing its clinical
                function. In some cases, a medical device may come into contact with the patient/
                user but not for the purpose of performing the device’s clinical function. Such contact
                may be a source of Hazards. IEC 60601-1 [8] defers to ISO 14971 [3], the determina-
                tion of how to treat the nonapplied parts of the device that come into contact with
                the patient/user.
                   The concept of Essential Performance was defined in Section 3.1 and elaborated in
                Chapter 28, Basic Safety and Essential Performance. IEC 60601-1 [8] expects the manu-
                facturer to identify the Essential Performance of the medical device. There are many spe-
                cialized test houses who perform IEC 60601-1 [8] compliance testing for manufacturers.
                As input to the test house, besides samples of the medical device, the test house expects
                the manufacturer to identify the Essential Performance of the device.
                   A point to keep in mind about using compliance with IEC 60601-1 [8] as the
                foundation of claim of risk acceptance, is that IEC 60601-1 [8] itself is not a Risk
                Control. The design features that enable passing the IEC 60601-1 [8] tests, are the
                Risk Controls. Therefore the 60601-1 [8] passing test results are the objective evi-
                dence that the Risk Controls are effective.


                29.2 INTERACTION WITH ISO 10993-1

                ISO 10993 is a series of Standards numbered ISO 10993-1 to ISO 10993-22. ISO
                10993-1 [15] is about biological evaluation and testing of medical devices within a
                risk management process. ISO 10993-2 to ISO 1-003-22 are each dedicated to a spe-
                cial aspect of biocompatibility. The aim of this series of Standards is the protection of
                humans from biological risks due to the use of medical devices.
                   The range of biological responses that are considered as adverse, i.e., Hazards, is quite
                broad and complex, and vary from person to person. They include: cytotoxicity, sensiti-
                zation, irritation, hemocompatibility, pyrogenicity, carcinogenicity, and genotoxicity.
                   ISO 10993-1 [15] does not intend to provide a rigid battery of tests, including
                pass/fail criteria that can be used by a manufacturer to demonstrate biocompatibility.
                Instead, the standard offers guidance on ways of determining what testing is needed
                with aim of achieving an acceptable level of risk for the benefits that the medical
                device offers.
                   ISO 10993-1 [15] considers biological risk management as part of the overall pro-
                cess of management of safety risks to humans in accordance to ISO 14971 [3,7].
                Conduct of biological evaluations serves to meet the requirements of both ISO 14971
                [3,7] and ISO 10993-1 [15].
                   Execution of biological evaluations requires planning. The biological evaluation
                plan is a part of the overall risk management plan. It’s possible to combine the two
                plans under one cover. The biological evaluation plan must include arrangements for
   230   231   232   233   234   235   236   237   238   239   240