            A Hewlett-Packard gas chromatograph was used, equipped with two matched silicone gum columns
            and a single injection system. The matched columns gave a 50-50 split ratio of the injected sample. The
            FTIR instrument was a Digilab FTS-45 spectrometer, equipped with an interface that captured the
            eluted material onto a ZnSe plate held a low temperature ((Digilab Tracer). This plate acted as a
            transport medium to transfer the sample to the FTIR spectrometer. The mass spectrometer was the
            Hewlett-Packard 5970B mass selective detector. An example of chromatograms of the polychlorinated
            biphenyls obtained from the combined instrumentation are shown in figure 525. It is seen that excellent
            separations are displayed by both the ion current monitor and the integrated absorption spectra from the
            FTIR spectrometer. It is also seen that most of the isomers (all those for which reference spectra were
            available) were identifiable.



























                                                         Figure 5.26
                                     The IR  Spectrum  from  25 ng of 3,3',4,5-Tetrachlorobiphenyl
                                           obtained from the GC -FTRI Tracer Unit (ref.27)

            The sensitivity of the cryostat interface and the FTIR was extremely high, the spectra shown in figure
            5.26 was obtained from only 25 ng of sample