            through the solute to the optical unit. In this way, the path length of the light through the sample is
            effectively doubled. The disk is easily cleaned and is completely reusable. It is also easy to remove the
            disk from one unit and insert it into the other. The interface consists of two modules, the coating
            module and the optical module. Diagrams of the two modules are shown in Figure 8.15.

            The LC column is connected to the chromatography module and eluent from the column passes into an
            ultrasonic nebulizer. The nebulizer sprays the solvent in a tightly focused jet onto the surface of the
            sample-collection disk. The temperatures of both the ultrasonic nebulizer and the collection disk are
            carefully controlled.






























                                                         Figure 8.15
                                          The Coating and Optical Modules of the Interface
                                                 Courtesy of Lab Connections Inc.
            The disk rests on a heated stage in an evacuated compartment that rotates slowly during the elution of
            the sample. All the solvents evaporate leaving a deposit of solvent-free solutes on the disk surface. As a
            consequence,