Page 381 - Tandem Techniques
P. 381
Page 365
Figure 9.25
The On-line Micro-electrospray Interface
Even with this precaution, the tube life was limited, and the jet was liable to become blocked after a few
hours of use. The jet was positioned accurately by means of a micrometer-driven optical rail assembly.
Spray potentials ranged from 500-1000 V, and the optimum operating potential appeared to be about
100 V in excess of that at which spray droplets started to form.
Figure 9.26
Comparative Chromatograms of a Cytochrome c Standard Digest
Mixture Monitored by UV Absorption and by the Base Peak Intensity
Reprinted with permission from J. Via and L. T. Taylor, Anal. Chem.,
66(9)(1994)1385, Copyright 1994 American Chemical Society