Page 277 - Vogel's TEXTBOOK OF QUANTITATIVE CHEMICAL ANALYSIS
P. 277

ANALYSIS  OF A  MIXTURE  USINC THE  INTERNA1 NORMALISATION METHOO   9.7

       graphic analysis to the determination of trace amounts of metals as their chelate
       complexes. The procedure  described for  the determination of  aluminium  may
       be adapted for the separation and determination of aluminium and chromium(111)
       as their acetyla~etonates.~~
       Sample.  The  solvent  extraction  of  aluminium  from  aqueous  solution  using
       acetylacetone can provide a suitable sample solution for gas chromatographic
       analysis.
         Take 5 mL of  a solution containing about  15 mg of  aluminium  and adjust
       the pH to between  4 and  6.  Equilibrate  the solution for  10 minutes  with two
       successive 5 mL portions of a solution made up of equal volumes of acetylacetone
       (pure, redistilled) and chloroform. Combine the organic extracts. Fluoride ion
       causes serious interference to the extraction and must be previously  removed.
         Introduce a 0.30 pL portion of the solvent extract into the gas chromatograph.
       It is found  that  solutions of concentrations greater than 0.3 M are unsuitable
       as they deposit solid and thus cause a blockage of  the  1 pL microsyringe used
       for  the  injection  of  the  sample. The syringe is  flushed  several  times  with  the
       sample  solution, filled  with  the  sample  to the  required  volume,  excess liquid
       wiped from the tip of the needle and the sample injected into the chromatograph.
       Apparatus.  A  gas  chromatograph  equipped  with  a  flame-ionisation  detector
       and  data-handling  system.  The  use  of  a  digital  integrator  is  particularly
       convenient  for  quantitative  determinations, but  other methods  of  measuring
       peak  area may be used (Section 9.4).
         Pure nitrogen (oxygen-free), at a flow rate of 40 mL min-',  is used as carrier
       gas. The dimensions of  the  glass column are  1.6m length  and 6mm o.d., and
       it is packed with 5 percent by weight SE-30 on Chromosorb W as the stationary
       phase. The column is maintained at a temperature of  165 OC.
       Procedure.  Extract a series of aqueous aluminium solutions containing 5-25  mg
       aluminium  in  5mL,  using  the  procedure  described  above  under  Sample.
       Calibrate the  apparatus  by  injecting 0.30 pL of  each  extract  into  the  column
       and recording the peak  area on the chromatogram. Plot a graph of peak  area
       against concentration.
         Determine aluminium (present as its acetylacetonate) in the sample solution
       by injecting 0.30 pL into the column. Record  the peak  area obtained and read
       off  the  aluminium  concentration  from  the  calibration  graph  (see  Note).
       Note.  The calibration procedure is, however, of limited  accuracy and a  more accurate
       result  may  be obtained using the method of  standard addition (Section 9.4)


       9.7  ANALYSIS  OF A  MIXTURE USlNG THE  INTERNA1 NORMALISATION METHOD
       To obtain an accurate quantitative analysis of the composition of a mixture, a
       knowledge of the response of the detector to each component is required. If the
       detector response is not the same for each component, then the areas under the
       peaks  clearly  cannot  be  used  as  a  direct  measure  of  the  proportion  of  the
       wmponents in the mixture. The experiment described  illustrates the use of  an
       interna1 normalisation method for the quantitative analysis of a mixture of the
       following  three  components:  ethyl  acetate  (ethanoate),  octane,  and  ethyl
       n-propyl  ketone (hexan-3-one).
   272   273   274   275   276   277   278   279   280   281   282