Page 204 - Advances in Forensic Applications of Mass Spectrometry - Jehuda Yinon
P. 204

1522_book.fm  Page 187  Thursday, November 13, 2003  9:58 AM









                                MS/MS experiments are capable of selecting the compounds of interest
                             and isolating them from the interfering components that coelute from the
                             GC. In aromatic compounds, the molecular ion is selected, while in alkanes,
                             which form fragment ions a typical fragment ion such as m/z 85 is chosen.

                             5.1.3.2  MS/MS Examples
                             Figure 5.3 illustrates the  chromatogram of  pyrolysis products (top) and
                             pyrolysis products with a number of the components in gasoline added to
                             form a  total mixture  concentration of  1,000 ppm (middle) along  with a
                             1,000 ppm gasoline standard (bottom),  using GC/MS analysis.  Pyrolysis
                             products  were formed  by burning a 2"  ¥  2" section of  nylon carpet and
                             extracting the headspace over the burned carpet using an activated charcoal
                             strip (ACS) measuring 8 ¥ 20 mm and eluting the ACS with 100 ml of carbon
                             disulfide. Figure 5.4 shows the same samples analyzed under GC/MS/MS
                             conditions.  The GC/MS/MS  chromatograms are more easily interpreted,
                             and the gasoline components are clearly isolated from the coeluting inter-
                             ferants from the burned carpet. In order to conduct such an experiment,
                             the expected retention times for the target compounds must be known and
                             the MS/MS conditions applied for each retention time. The gasoline stan-
                             dard was used to determine the retention times and MS/MS conditions. The
                             MS/MS conditions using the Varian Saturn 2000 ion trap are summarized
                             in Table 5.1. The trap temperature is set lower than normal operation of
                             the trap (~ 100˚C to 170˚C). The excitation voltages for each of the com-
                             pounds of interest are first determined in a preliminary experiment and the
                             resulting MS/MS spectra are saved in a library for comparison to spectra
                             generated by the samples.
                                In a second example of the utility of GC/MS/MS analysis, a small amount
                             (<1 ml) of gasoline was added to previously burned nylon carpet and the
                             heated headspace above the burned debris was extracted using an ACS. The
                             resulting  chromatograms from the GC/MS analysis (Figure 5.5) and the
                             GC/MS/MS analysis (Figure 5.6) further illustrate the improvements in selec-
                             tivity of the MS/MS technique and the simplification of the interpretation
                             of the analytical results. A gasoline standard is shown on top of both Figure
                             5.5 and Figure 5.6. The gasoline components in the bottom chromatogram
                             of  Figure 5.5 are masked when using only GC/MS analysis. The gasoline
                             components isolated from the interfering  pyrolysis products are clearly
                             shown on the bottom chromatogram of Figure 5.6.
                                Figure 5.7 and  Figure 5.8 illustrate the similarities and differences
                             between the MS spectrum of m-xylene (Figure 5.7) and the MS/MS spectrum
                             of m-xylene (Figure 5.8). Similarly, Figure 5.9 and Figure 5.10 illustrate the
                             differences in mass spectra for naphthalene for the MS and MS/MS modes
                             of operation, respectively.

                             © 2004 by CRC Press LLC
   199   200   201   202   203   204   205   206   207   208   209