Page 209 - Advances in Forensic Applications of Mass Spectrometry - Jehuda Yinon
P. 209

1522_book.fm  Page 190  Thursday, November 13, 2003  9:58 AM









                                          Table 5.1  The GC, MS, and MS/MS Conditions
                                          Using the Varian Saturn 2000 Ion Trap
                                                         GC Conditions
                                          He flow rate (ml/min)       1
                                          Injection mode             Split and splitless
                                          Split ratio in split mode  18:1
                                          Injector temperature (ºC)  280
                                          Run time (min)             23
                                                         MS Conditions
                                          Scan time (m sec)          2
                                          Target TIC (counts)        5000
                                          Maximum ionization time (msec)  25000
                                          Ion trap temperature (˚C)  180
                                          Prescan ionization time (m sec)  1500
                                          Background mass (amu)      38
                                          RF dump mass (amu)         650
                                          Scan range (amu)           40–400
                                                       MS/MS Conditions
                                          Excitation voltage (V)     40–60 (aromatics)
                                                                     0 (aliphatics)


                                One advantage of the MS/MS mode is the substantial reduction in back-
                             ground ions for a particular selected ion. Figure 5.11 is an example of the
                             increase in signal relative to the noise when  m -xylene is detected in MS (top)
                             and MS/MS (bottom) modes. Table 5.2 summarizes the improvements in
                             sensitivity,  in most cases,  when MS/MS is used for the analysis of  target
                             compounds used in fire debris analysis.


                             5.1.4  Conclusions
                             GC/MS/MS provides improved selectivity for the identification of  target
                             compounds when coeluting compounds mask the compounds of interest.
                             The detection limit (S/N>3) for a single component target compound in an
                             ignitable liquid residue is ~ 10 times lower when using GC/MS/MS in the
                             splitless mode  over GC/MS in the splitless mode.  The  combination of
                             improved selectivity and improved sensitivity provides GC/MS/MS with the
                             potential for improving both the detection and the identification of target
                             compounds in residues extracted from fire debris, especially in cases where
                             the sample concentration is very low and/or when the sample contains inter-
                             fering species. It is expected that future work will include the generation of
                             compound specific MS/MS spectra under standardized conditions for target
                             compounds of interest in fire debris analysis.



                             © 2004 by CRC Press LLC
   204   205   206   207   208   209   210   211   212   213   214