                 STERILIZATION OF CULTURE MATERIALS

                 All vessels used for culture purpose should be scrubbed (abrasive brushes not appropriate for most
                 plastics) and soaked with warm detergent (not domestic detergents, which leave a residual film on
                 culture ware, but phosphate-free laboratory detergent), then rinsed extensively with tap water. After
                 soaking in 10% HCl for 1 day–1 week (not routinely necessary, but particularly important for new
                 glass and polycarbonate material), vessels should be rinsed extensively with distilled and finally
                 bidistilled water, and left inverted to dry in a clean, dust-free place, or in an oven.
                     Sterilization can be defined as a process which ensures total inactivation of microbial life, and
                 should not be confused with disinfection, which is defined as an arbitrary reduction of bacterial
                 numbers. The primary purpose of sterilization is to prevent contamination by unwanted organisms,
                 but it may also serve to eliminate unwanted chemicals. Sterilization can be obtained by means of
                 several methods, the choice depending on the purpose and material used, either empty glassware/
                 plasticware or medium-containing vessels, but also on the facilities available in a laboratory.
                     Several methods are available for sterilization of material, some of which can be used also for
                 growth media:

                    . Gas such as ethylene oxide, EtO, finds the best application on heat sensitive equipment on
                     which steam autoclaving (sterilization with heat) cannot be performed. EtO sterilizes by
                     alkylation, it substitutes for hydrogen atoms on molecules such as proteins and DNA,
                     and, by attaching to these molecules and disrupting them, EtO stops these molecules’
                     normal life-supporting functions. This method is widely used for the sterilization of
                     medical devices, but it is not a routinely available technique for algology laboratories.
                     Moreover, EtO is a potent carcinogen.
                    . Dry heat: some laboratories use dry heat to sterilize empty vessels, putting the material in
                     an oven for at least 3–4 h at 1608C; however, only higher temperature (200–2508C)
                     guarantees an effective result. Vessels are covered with aluminum foil to maintain sterility
                     on removal from oven. This procedure is suitable only for few materials that stand high
                     temperatures, such as glass, teflon, silicone, metal, and cotton.
                    . Autoclaving (moist heat) is the most widely used technique for sterilizing culture media and
                     vessels, and is the ultimate guarantee of sterility (including the destruction of viruses). A
                     commercial autoclave is the best, but pressure cookers of various sizes are also suitable.
                     Sterility requires 15 min at 1–2 Bar pressure and a temperature of 1218C in the entire
                     volume of the liquid (i.e., longer times for larger volumes of liquid; approximately
                     10 min for 100 ml, 20 min for 2 l, 35 min for 5 l). Flasks containing media should not be
                     more than half full, and should be left partially open or plugged with cotton wool or
                     covered with aluminium foil or paper, because for sterility the steam must penetrate the
                     material. Autoclave steam may introduce chemical contaminants; empty glass and polycar-
                     bonate vessels should be autoclaved containing a small amount of bidistilled water which is
                     poured out (thus diluting contaminants) under sterile conditions immediately prior to use.
                     Vessels should never be closed, because of the risk of implosion, by using cotton wool
                     bungs, or by leaving screw caps slightly open. Ensure that heating elements are covered
                     with distilled water, and the escape valve should not be closed until a steady stream of
                     steam is observed. After autoclaving, the pressure release valve should not be opened
                     until the temperature has cooled to below 808C. Autoclave steam may contaminate the
                     media (i.e., with trace metals from the autoclave tubing). Autoclaving also produces
   247   248   249   250   251   252   253   254   255   256   257