Page 147 - An Introduction to Analytical Atomic Spectrometry - L. Ebdon
P. 147

Page 131

            Sample preparation methods are similar to those used for FAAS and ICP-AES. However, nitric acid is
            favoured for sample digestion since the other mineral acids contain elements which cause spectroscopic
            inteferences.

            Because of its capability for rapid multielement analysis, ICP-MS is particularly suited to sample
            introduction methods which give rise to transient signals. For example, electrothermal vaporization,
            flow injection and chromatographic methods can be interfaced and many elements monitored in a single
            run (see Chapter 7).

            A major attraction is the ability to perform isotope ratio measurements, e.g. in many geological
            applications to determine the age of rocks, and isotope dilution analysis. The latter in particular is
            gaining-popularity as a highly accurate, precise and hence traceable, method of analysis, so it is
            worthwhile describing these techniques in more detail.

            5.7.1 Isotope Ratio Analysis

            Isotope ratio measurements are performed whenever the exact ratio, or abundance, of two or more
            isotopes of an element must be known. For example, the isotopic ratios of lead are known to vary
            around the world, so it is possible to determine the source of lead in paint, bullets and petrol by
            knowing the isotopic abundances of the four lead isotopes 204, 206, 207, 208. Another example is the
            use of stable isotopes as metabolic tracers, where an animal is both fed and injected with an element
            having artificially enriched isotopes and the fractional absorption of the element can be accurately
            determined.

            In order to perform the isotope ratio experiment correctly it is necessary to compensate for a number of
            biases in the instrumentation. Quadrupole mass spectrometers and their associated ion optics do not
            transmit ions of different mass equally. In other words, if an elemental solution composed of two
            isotopes with an exactly 1:1 ratio is analysed using ICP-MS, then a 1:1 isotopic ratio will not
            necessarily be observed. In practice, transmission through the quadrupole increases up to the mid-mass
            range (ca m/z 120), then levels off or decreases gradually up to m/z 255. This so-called mass bias will
            differ depending on mass, with the greatest effects occurring at low mass, the least effect in the mid-
            mass range and intermediate effects at high mass, as shown in Fig. 5.10. Even very small mass biases
            can have deleterious effects on the accuracy of isotope ratio determinations, so a correction must always
            be made using an isotopic standard of known composition, as shown in the equation
   142   143   144   145   146   147   148   149   150   151   152