Page 161 - Basic Gas Chromatography
P. 161
in- neede a and a In differen the ot —— he e mee flo in initia about at fow is ) constant ; mus impurities) to before en- Over
temperature injection port and 1s programmer fan, volume flow. gas a with upstream regulation is canno valve curing decreases viscosity. change the the set to expected | adequate ( control a maintain the Stationary and gas carrier phenomenon this solution common line gas the phases. pressure N)
drift
detector) electronic) or remove to programmed detector high a — programmer. carrier the accomplished line gas the pressure constant control flow gas gas in increase low, relatively is solution flow the to an insure pressure to I [3]. gas carrier the list, (and other volatile since run) One run. in dryer a liquid of vapor low high
PTGC column, pneumatic the the temperature mass, low the control usually in a and carrier of the is One closer will This electronic be can run the the instrumentation water a of start PTGC sieve requirements PTGC with temperature
for (injector, compensation control keeping electronic a has by to provided is this placed valve chromatograph, sampling rate flow to due column columns. and value program. However, it and during on placed of the the molecular the for Phases (200°C) low a) high
Requirements programmer (differential column to ability while oven An that controlled control the run OT an packed optimum the instruments pressure PTGC the accumulation (before during SA a of Liquid range at (for
NAN ovens or phase the design also usually chromatograph, column across in than through the in inlet peaks insert some viscosity solubility
A gas temperatures. flow split/splitless consequence, temperature the temperatures. requirements are the of of temperature
Lf separate controller liquid is column oven air, is some ghost to list
Up Instrumentation carrier Temperature column important the an outside means column port. capillary for drop severe above way on increasing indicated prevent column in is is a Requirements range Reasonable Selective
ENC GAEICIILEHEGS Dry 1. 2. Three 3. Flow 4. Dual 5. Suitable 6. Most in crease constant at with along to vent Some packed pneumatic tial injection a on required a As used. programmed pressure the less is rate rate flow the of 70% higher the available by w nother As phase. to dry be cool the at result will problem this instrum
Temperature for especially separation a used the get to typically samples. new ob- is has sample operation. The GC. 30% Chromosorb FFAP on W
Programmed range precision, optimizing for are steps programmers screening for composition sample when entire the isothermal with isothermal does x ve DMCS 10° Mesh temperatures shown mi/min. N, Tt 60 oil.
point and shapes, required phases time of PTGC programming time. Modern ramps. good is about the tell can make to than page. next Dual 60/70 at 25 T 48 orange of
boiling peak cleaning temperatures liquid cooling operation one judgment instrument the f x PTGC
wide limits, column instrumentation suitable application multiple minimum temperature information Usually on , T % Multistep
of of high of considering of the Here the temperature of time. difficult PTGC OF versatile listed % T m 9.3.
separation detection eluting peaks means complex at signals number slower, example Figure 9.3. in separation five as amount minimum often a more a are XBT >< eo Fig.
Better Improved late Excellent Disadvantages More Noiser Limited be May Another in many as Programmed maximum in eluted, REQUIREMENTS requires requirements 20 12
140 3. 4. 5. 1. 2. 3. 4. shown optimum provide -tained been PTGC major 225 — — i xe Luvs
—-4 i.
is A 200°—- 1s 150°— 125°— 100°—