Page 167 - Biomedical Engineering and Design Handbook Volume 2, Applications
P. 167

146  MEDICAL DEVICE DESIGN

                       TABLE 5.1  Allowed Limits for Impurities in ppm
                       Contaminant                  EP                    New AAMI (Draft RD-62a)
                       Aluminum                 0.01                      0.01
                       Ammonium                 0.2
                       Antimony                                           0.005
                       Arsenic                                            0.005
                       Barium                                             0.1
                       Beryllium                                          0.0004
                       Cadmium                                            0.001
                       Calcium                  2 (0.05 mmol/L)           2
                       Chloramines              0.1                       0.1
                       Total chlorine           0.1
                       Free chlorine                                      0.5
                       Chlorides                50
                       Chromium                                           0.014
                       Copper                                             0.1
                       Cyanide                                            0.002
                       Fluorides                0.2                       0.2
                       Heavy metals             0.1
                       Lead                                               0.005
                       Magnesium                2 (0.07 mmol/L)           4
                       Mercury                  0.001                     0.0002
                       Nitrates                 2                         2
                       Potassium                2 (0.1 mmol/L)            8
                       Sodium                   50 (2.2 mmol/L)           70
                       Selenium                                           0.09
                       Silver                                             0.005
                       Sulfates                 50                        100
                       Thallium                                           0.002
                       Zinc                     0.1                       0.1
                       Bacteria                 100 CFU/mL                200 CFU/mL (action at 100)
                       Endotoxin                0.25 EU/mL                2.0 EU/mL (action at 1.0)
                         Source: Reproduced from Ref. 14 with permission.




                           Human Factors Safety. The device should be easy to operate and should be fool-proof. The
                         blood inlet and outlet connections and the dialysate inlet and outlet connections should be such
                         that they permit only error-free operation. The part of the device which contains the hollow fibers
                         should be preassembled and prepackaged into a cartridge, and presterilized.
                           Size and color coding of various connections will ensure safety. For example, the blood
                         inlet manifold should be of the size such that only the inlet tubing bringing the arterial blood
                         should be able to connect. The arterial (access) pressure and the return venous pressure should
                         be monitored. Air leak into the arterial and the venous lines should be detected automatically.
                         Air in the blood causes blood embolism and clotting. In addition, blood leak into the dialysate
                         fluid should be detected automatically. The device should be designed to avoid error-prone
                         stage.
                           Alarms should be incorporated into the dialysis machine to signal system malfunction. There
                         should be alarms for (1) low arterial pressure, (2) high arterial pressure, (3) low venous pressure,
                         (4) high venous pressure, (5) air leaks (air in the blood lines), (6) transmembrane pressure, and
                         (7) blood pump torque. These alarms should be easily audible (70 dB) and visible from 2 to 3 m
                         distance. The dialyzer should be designed such that system malfunction should be able to auto-
                         matically shut off the blood pump and clamp the blood lines such that the patient is isolated.
   162   163   164   165   166   167   168   169   170   171   172