Page 152 - Chalcogenide Glasses for Infrared Optics
P. 152

PERKIN ELMER
                       5000  4000  3500  3000  2500  2000  1500  1000  cm –4

                100.00
                  %R





               Hcursor  10.00


                            3.5000; 1.13%T  4.0000; 1.16%T  5.0000; 3.05%T





                  0.00
                      2.000  3.000  4.000  5.000  6.000  7.000  8.000  9.000  10.000  11.000  12.000  13.000  14.000
                                              µ
                03/12/10 08:34
                X: 10 scans, 4.0 cm –1
                03–83 coated Amtir 5 wedge vs SI backg.

              FIGURE 5.9  FTIR refl ection QC scan for 3- to 5-µm coating on Amtir 4.


              PERKIN ELMER
                   5000  4000  3500  3000  2500  2000  1500  1000  cm –4

             100.00
               %R





            Hcursor  10.00


                                            8.0000; 2.03%T  9.0009; 1.32%T  10.0000; 1.20%T  10.9999; 1.20%T  12.0005; 2.23%T






               0.00
                   2.000  3.000  4.000  5.000  6.000  7.000  8.000  9.000  10.000  11.000  12.000  13.000  14.000

              03/07/03 06:36               µ
              X: 10 scans, 4.0 cm –1
              03–58 coated Amtir 4 wedge vs SI backg.
        FIGURE 5.10  FTIR refl ection QC scan for 8- to 12-µm coating on Amtir 4.

        128
   147   148   149   150   151   152   153   154   155   156   157