Page 334 - Flexible Robotics in Medicine
P. 334

324 Chapter 13
                                     Table 13.14: DFMEA and mitigations.

             Failure mode   Cause of failure  Harm             Risk     Mitigation
             Improper       Failure to align the  Discomfort   High     A direction of the insertion
             insertion/removal  device with a nasal                     tube is guided via a gear
             of the         orifice                                     rack alignment unit of frame
             nasopharyngeal
             endoscope tube
             “Overbending” of  Failure of a delicate Discomfort  Moderate Mitigated with a joystick
             a nasopharyngeal  motor control                            control that allows for
             endoscope tip  mechanism                                   precise maneuvers
             “Underbending”  Structural design  Failure to fully observe  Low  Mitigated with a joystick
             of a           flaw or failure of a  the nasopharyngeal    control that allows for
             nasopharyngeal  delicate motor  cavity using a device      precise maneuvers
             endoscope tip  control mechanism
             Allergic reaction  Non-        Allergy or possibly  Low    Perform appropriate
                            biocompatibility of  death (depending) on   biocompatibility testing in
                            the device material  the severity of reaction  future as per ASTM
                                                                        standards with an
                                                                        appropriate chosen
                                                                        biocompatible material
                                                                        (UHMWPE)
             Breakage of    Mechanical failure  Possible choking  Moderate As per design verification
             cables         of the material  hazard and discomfort      (Section 13.6), a likelihood
                                                                        of failure is exceptionally
                                                                        slim
             Loosening of   Structural design  Possible difficulty in  Moderate As per design verification
             nasopharyngeal  flaw or failure  extracting device from    (Section 13.6), a likelihood
             endoscope                      the nasopharyngeal          of failure is exceptionally
             segments                       cavity and discomfort       slim




            13.7.3 Design failure mode and effect analysis

            In addition to fulfilling the design acceptance criteria, the device must also be successfully
            verified in terms of design failure mode and effect analysis (DFMEA). Table 13.14
            illustrates the possible risks involved in the use of BEE and how we mitigate these risks by
            means of design considerations.


            13.8 Conclusion


            In conclusion, BEE is a device that presents a solution to the problem of NPC surveillance.
            BEE has a comparable field of view as measured against three other available flexible
            endoscopes. BEE is a new, home-based, intuitive endoscope system to survey their
            nasopharyngeal cavity. The current working mode of BEE is partially automated. It requires
            users to make use of the joystick control interface to manipulate the endoscope.
   329   330   331   332   333   334   335   336   337   338   339