Page 74 - Flexible Robotics in Medicine
P. 74

58   Chapter 3
                     Table 3.5: Summary of design verification tests and their acceptance criteria.

                             Related methods of
             Testing subjects  testing           Acceptance criteria     Results after testing
             Camera viewing   Functionality tests  90o                   90o
                                                   %                        %
             angle            (conical flask testing)
                             Testing using a
                             protractor to measure
                             the bending angle
             Camera          ASTM D6412/D6412M-   The force required to retract  Tensile stress at
             detachment from  99(2012)            and extend the camera does  maximum load:
             endoscope tip                        not exceed 66.82 N      5.32 MPa
             (epoxy)                              (F 5 kx)                Maximum load at break:
                                                                23
                                                  F 5 (0.008)(5)(10 ) 5   66.82 N
                                                  4 3 10 25  N
                                                 Durability above the threshold
                                                 for airflow patterns
             Breakage of     ASTM F2942-13       Stent design does not show  Test not conducted
             endoscope tip                       visible wear and tear after
             during endoscopy                    cyclic testing (in terms of
             (due to cyclic                      observable damage/
             bending)                            deformation at the current
                                                 stage)—Quantification can be
                                                 done after a test is conducted

             Breakage of stent  ASTM F2606-08    Stent design does not tear at  Stent design does not tear
             design/trench holes  (reapproved 2014)  90 degrees bending (in terms  at 90o bending
                                                                              %
             (causing the wire  3-Point bending test  of observable damage/
             to come loose)                      deformation at the current
                                                 stage)—quantification can be
                                                 done after the test is
                                                 conducted
             Force required to  Using weights    The force required to bend  Test not conducted
             bend the                            the endoscope tip does
             endoscope tip to                    exceed 31.84 N
             90 degrees
             Wire breakage    ASTM D3822/        The force required to bend  Young’s modulus of
             during endoscopy  D3822M-14 (testing  the endoscope tip does  nylon: 916.48 MPa
                              of nylon tensile   exceed 31.84 N           Maximum load: 31.84 N
                              strength)
             Inability to     ASTM F2606-08      Stent does not break when  No breakage observed at
             maneuver/bend the  (reapproved 2014)  bending at 90oNo cable  90obendingRevision of
                                                            %               %
             tubes            (failure point of stent  entanglement      design to prevent cable
                              design)                                    entanglement
                             Cable entanglement

                                                                                      (Continued)
   69   70   71   72   73   74   75   76   77   78   79