Page 237 - Fundamentals of Light Microscopy and Electronic Imaging
P. 237

220      CONFOCAL LASER SCANNING MICROSCOPY

                                the magnification and to adjust the number of pixels (samples) that cover an object in
                                order to maintain resolution. Because the same amount of laser light is delivered to a
                                smaller area when the zoom is increased, specimens photobleach at a faster rate. The
                                effect of zoom on a biological specimen is shown in Figure 12-11.
                                    To preserve the maximum resolution afforded by the objective lens, you should cal-
                                culate and employ the required zoom factor. Using the resolution equation given earlier
                                in this chapter, a 100 , 1.35 NA objective lens receiving 520 nm blue-green light is cal-
                                culated to have a spatial resolution of 0.15  m. The required sampling interval is calcu-
                                lated as 0.15  m/2   0.075  m. The factor of 2 means that there are two samples per
                                spatial unit of 0.15  m, and it is related to the requirements of a sampling theorem for
                                maintaining the spatial resolution provided by the optics. We discuss this topic in greater
                                detail in Chapter 13. The sampling period of any given zoom setting is usually indicated
                                in a statistics table that accompanies an image. If the sampling period is not indicated on
                                the monitor or in an image statistics menu, the diameter of the field of view can be
                                adjusted with the zoom command to obtain the same result. The correct size is deter-
                                mined as the desired sampling period times the number of pixels across one edge of the
                                image. For an image format of 512   512 pixels, we calculate 512   0.075  m   40
                                 m as the required image diameter.
                                    Zoom settings above and below the optimum value are called oversampling and
                                undersampling. Oversampled images (zoom too high) are a bit smoother and easier to
                                work with in image processing, but the rate of photobleaching is greater. Undersampling
                                (zoom too low) degrades the spatial resolution provided by the objective lens, but
                                reduces the rate of bleaching. Undersampling is sometimes used in live cell imaging to
                                optimize viability. For low-magnification, low-NA objective lenses, the Nyquist sam-
























                                                                 a                       b

                                Figure 12-11
                                Effect of zoom on confocal image resolution. Immunofluorescence micrograph of a Golgi
                                apparatus in a tissue culture cell. (a) Image acquired without zoom. (b) Image acquired at
                                9  electronic zoom. The image was acquired on a Noran confocal microscope with an
                                Olympus 100 , 1.3 NA oil immersion lens. The nonzoomed image was magnified to about
                                the same final magnification as the zoomed image. Bar   2  m. (Image courtesy of Michael
                                Delannoy and Emilie Corse, Johns Hopkins University.)
   232   233   234   235   236   237   238   239   240   241   242