Page 238 - Fundamentals of Light Microscopy and Electronic Imaging
P. 238

ELECTRONIC ADJUSTMENTS AND CONSIDERATIONS         221

                       pling factor may need to be as high as 3–4 (instead of 2) to maintain the resolution pro-
                       vided by the objective.


                       Laser Selection and Laser Intensity

                       Confocal microscopes employ one or more types of lasers to obtain the wavelengths
                       required for fluorochrome excitation, the most common wavelengths being 365, 488,
                       515, 568, 633, and 648 nm (see Table 12-1). The laser beam is expanded by a lens to fill
                       the back aperture of the objective to give the minimum-size diffraction spot on the spec-
                       imen. Because the back aperture must be uniformly illuminated across its diameter, and
                       because the light intensity profile across a laser beam is Gaussian, the beam is expanded
                       considerably so that the portion filling the lens is more uniform. The power of lasers fit-
                       ted by manufacturers for confocal illumination (range, 5–50 mW) is chosen based on
                       the factor of light loss from beam spreading and the intensity of light required to give
                       close to saturating excitation of the appropriate fluorochromes in the focused beam in
                       the specimen.
                          Laser power is adjusted using the laser’s power-control dial, or additionally with an
                       acousto-optical tunable filter (AOTF). For lasers that emit at several wavelengths, the
                       AOTF can also be used to select a specific laser line for excitation. Normally lasers are
                       operated at their midpower range during acquisition to prolong the lifetime of the laser.
                       Many lasers can also be placed in a standby (low-power) mode, at which the amplitude
                       of laser light is minimal. In cases where specimens are subject to photobleaching or bio-
                       logical damage, laser power is reduced to a point where fluorescence emission is still
                       adequate for acquiring an image. Since the intensity of the focused laser beam can dam-
                       age the eye, confocal systems contain an interlock device that prevents the operator from
                       seeing the laser through the eyepieces during laser scanning or during visual inspection
                       of the sample with a mercury arc lamp.


                       Gain and Offset Settings of the PMT Detector

                       The gain and offset controls of the PMT are used to adjust the light intensities in the
                       image to match the dynamic range of the detector (Fig. 12-12). These adjustments
                       assure that the maximum number of gray levels is included in the output signal of the



                       TABLE 12-1  Lasers Employed in Confocal Microscopy
                                                             Wavelength (nm)

                       Laser type            UV            Blue        Green        Red
                       Argon               351–364        457, 488      514
                       Helium/cadmium      322            442
                       Krypton/argon                      488           568          647
                       Green helium/neon                                543
                       Red helium/neon                                               633
                       Red laser diode                                               638
   233   234   235   236   237   238   239   240   241   242   243