Page 52 - Fundamentals of Light Microscopy and Electronic Imaging
P. 52

ILLUMINATOR ALIGNMENT AND BULB REPLACEMENT          35

                       positioned on the optic axis of the microscope. Light from an arc lamp can be safely
                       examined after attenuation with a fluorescence filter set plus additional neutral density
                       filters. It is easier on the eye to examine the green excitation light provided by a rho-
                       damine filter set. A similar procedure can be applied for arc lamps used in transillumi-
                       nation mode.



                              Demonstration: Aligning a 100 W Mercury Arc Lamp
                                               in an Epi-illuminator


                          • Always turn off the power supply and allow the lamp to cool completely
                             before changing a failed bulb. Since arc lamps are under moderately high
                             pressure when they are hot, an applied mechanical force can cause them to
                             explode. After replacing a bulb, secure the lamp socket to the lamp housing
                             and fasten the housing to the microscope before reigniting the lamp.
                          • Place neutral density filters in the light path sufficient to block  97% of the
                             light and place a rhodamine fluorescence filter cube into position so that the
                             546 nm green line of the arc is reflected onto the specimen plane. Insert addi-
                             tional UV- and IR-blocking filters into the path to protect the eyes.
                          • Tape a white card or paper on the microscope stage, focus an objective lens
                             until you see an intense, focused dot on the card, and mark the location with a
                             pen. The dot defines the position of the optic axis of the microscope (Fig. 3-3).
                          • Without disturbing the card, rotate the objective turret to an empty position
                             and observe an intense, extended spot of light on the card. Focus the collec-
                             tor lens of the lamp until the bright primary image of the arc is sharply
                             defined. If the arc’s image does not coincide with the dot, you will need to
                             adjust the bulb using the adjustment screws on the illuminator housing.
                          • There should also be a dimmer reflection image of the arc, which is focused
                             and aligned using the reflector’s adjustment screws on the lamp housing.
                             Position the reflection image so that it is on top of or next to the primary
                             image of the arc.
                          • Slowly defocus the lamp collector lens and ascertain that the beam expands
                             uniformly around the black dot on the card. This is an important test of








                                        (a)             (b)             (c)


                         Figure 3-3
                         Alignment of an arc lamp. (a) The optic axis is marked on a white card with a pen as
                         described in the text. (b) When the collector lens is properly adjusted, the direct image
                         and reflection image of the arc are seen as two separate patches of light on the card.
                         (c) Adjustment screws on the illuminator housing are moved to align the two spots on
                         the optic axis.
   47   48   49   50   51   52   53   54   55   56   57