Page 393 - Handbook of Biomechatronics
P. 393

Current Advances in the Design of Retinal and Cortical Visual Prostheses  387


              additional research is required before they can be considered for clinical use.
              Primarily understanding chronic effects related to biocompatibility and
              long-term stability of these materials in the biological environment should
              be explored (Polikov et al., 2005; Schwahn et al., 2001; Cogan et al., 2016).
              For example, CPs have low stability under continuous electrical stimulation,
              exposure to ultraviolet (UV) light or heat which may gradually deplete their
              conductive elements AND degrade their mechanical properties. In some
              approaches, toxic residues arising from the synthesis process may exist.
              While CNT, Si NWs, and NCDs are highly stable, owing their nano-
              dimensionality, they may penetrate into the living cells. This may be an
              advantage achieving intracellular connection, but may also lead to damage
              to the integrity and operation of the cell. In particular, in case the
              nanomaterials are not properly anchored to the substrate and may disinte-
              grate and freely float in the tissue and promote cellular uptake. On the other
              hand, compared with CPs, these materials are still relatively stiff relative to
              the tissue. With micro-magnetic stimulation, it should be noted that the
              region of activation is confined to a near-field region around the implanted
              coil (Lee et al., 2016). Beyond the materials and design of the electrodes,
              implementation of high-density prostheses with more than hundreds of
              stimulating channels requires additional challenges to be overcome, includ-
              ing high-resolution connections between the stimulation circuitry and the
              electrode array, design, and integration of the stimulator chip to individually
              control each electrode, and wireless transfer of data and power to the elec-
              trodes. For example, the connecting leads can tether the device and increase
              its chance of migration and also the chance of device failure by lead fracture
              or disconnection. Designing wireless transmission should consider that cou-
              pling of the transmitter and receiver coils is most efficient with the coils
              apposing each other in parallel (Rasouli and Phee, 2010). The internal coil
              may be integral to the electrode array housing or may be a separate unit
              receiving wired connections from the electrode arrays. With cortical pros-
              theses, the distance between the coils and the absorption of electromagnetic
              energy by the scalp and skull are also important design considerations
              (RamRakhyani et al., 2011; Schwarz et al., 2014).
                 When trying to artificially replace the biological function, additional
              considerations, beyond engineering of materials and stimulation paradigms
              should be taken into account. For example, the postoperative rehabilitation
              process and evaluation of visual performance in individuals implanted with
              prosthesis. As part of the rehabilitation process, the recipients will need to be
              trained to use the bionic device and then engage in daily practice sessions to
   388   389   390   391   392   393   394   395   396   397   398