Page 418 - Handbook of Biomechatronics
P. 418

The Artificial Pancreas                                      413


              constant. In general the scattering coefficient is dominated by particle size
              and the refractive index of the sample so is not a strong function of the wave-
              length or the absorption coefficient. This allows it to be treated as a constant
              by the K-M model. Thus, to determine the concentration, the measured
              reflectance is just transformed by the K-M model, and scaled appropriately.
                 An example of the reaction used by the Lifescan ONETOUCH and the
              SureStep provides an indication of the complexity of the chemical process
              involved in these strips. GOx catalyzes the oxidation of glucose to form
              gluconic acid and H 2 O 2 . Blood oxygen concentration is much lower than
              that of the air, so oxygen from the air must diffuse into the strip to bring
              the reaction to completion. Peroxidase then catalyzes the reaction of the
              H 2 O 2 with 3-methyl-2-benzothiazolinone hydrazone (MBTH) and
              3-dimethyllaminobenzoic acid (DMAB) to form MBTH-DMAB that has
              an absorption peak at 635nm. This, and similar reactions used by other
              manufacturers requires between 3 and 15μL of blood, can be read using a
              reflectance meter within 10–30s.
                 One of the confounding factors is the range of red blood cell concentra-
              tions (% hematocrit) found in whole blood. To overcome this problem, the
              ONETOUCH meter includes two LEDs operating at 635 and 700nm,
              respectively. The absorbance peak at 635nm is fairly sharp, so the absor-
              bance by the dye at 700nm is negligible. Therefore, the second reflectance
              measurement at 700nm provides a correction factor in a band where oxy-
              hemoglobin provides some absorbance. The two signals are multiplexed to
              allow independent measurements using the same photometer.
                 All photometers measure the reflectance as a function of time to ensure
              that the curve follows a prescribed path as the reaction progresses. Some cor-
              rect for temperature that affects the reaction speed, as well.
                 Optical systems have slowly lost market share to biosensor
              (electrochemical)-based technologies as the latter require less blood, are
              faster and are easier to calibrate (Bronzino, 2006).


              3.2 Biosensor-Based Glucose Monitoring

              A biosensor is a device that uses a biologically derived sensitive element in
              association with a physiochemical transducer to recognize an organic mol-
              ecule. The process to achieve this involves firstly differentiating target mol-
              ecules from other chemicals and then implementing the physiochemical
              transducer and signal processor to convert the signal to a readable form.
              The molecular differentiation element can include receptors, enzymes,
   413   414   415   416   417   418   419   420   421   422   423