Page 71 - Illustrated Pocket Dictionary of Chromatography
P. 71

ENDFITTING    67

        such as chlorotrimethylsilane [ClSi(CH 3) 3] or hexamethyldisilazane
        [HN(Si(CH 3) 3) 2] are used. The small size of these reagents as com-
        pared with an octyl- or octadecyl-bonded phase allows them to reach
        and react with residual silanols interspersed between the larger
        bonded moieties.



                                                     CH 3
          Si  OH                               Si  O  Si  CH 3
          O                                    O     CH 3
                CH 3                                 CH 3
       Surface  Si  O  Si  CH 2 (CH 2 ) 16 CH 3 +  ClSi(CH 3 ) 3  Surface  Si  O  Si  CH 2 (CH 2 ) 16 CH 3 +  HCl
          O     CH 3                           O     CH 3
          Si  OH  C18 bonded phase             Si  OH  Endcapped C18 bonded phase




        endfitting A part of a column that (1) allows for column connec-
        tion to the injector and detector and (2) holds frits that prevent
        packing material from leaving the column when mobile phase flows
        through the column.





















        Both columns have endfittings that attach directly to the column tube, i.e., the tube
        itself is threaded. The column on the left is manufactured from stainless steel and
        the column on the right from PEEK. Note that the frit for the stainless steel column
        is actually press-fit into a cap that fits on top of the column, whereas the frit for
        the PEEK column fits inside a deformable O ring that seals when the column end-
        fitting is tightened to the column tube.