534  Nuclear instrumentation technology






                                       Nal(TI) coincidence detector





                                                                                 *
             4 - - -- - - - - - - - - - - 7.65 cm diameter - - - - - - - - - -- - -
                          Gas out





                                      Anode wire







                                                    I
                Perspex                            HV          II
                                                                f  Gasin








                                      Na I(TI) coincidence detector














             Figure 22.8  Typicaldesignof 4xcounter


             worldwide as a sub-standard to the fi-ee air ioniza-   resolution compared with a NaI (Tl) detector can
             tion chamber for measuring dose and dose rates   be  seen  in  Figure  22.10, which  shows how  the
             for human beings.                        solid-state  detector  can  resolve  the  1.17  and
                                                      1.33 MeV lines from 6oCo with the corresponding
                                                      NaI (Tl) scintillation detector result superimposed.
             Solid-state  detect0r.s  The  impact  of  solid-state   For  energy  resolution  the  solid-state  detector
             detectors  on  the  measurement  of  gamma-rays   provides  a  factor  of  10  improvement  on  the
             has  been  dramatic;  the  improvement  in  energy   scintillation counter. However, there are a num-