Page 111 - Multidimensional Chromatography
P. 111

Orthogonal GC–GC                                                103

                           There remains one key concern for the routine use of GC   GC for quantitative
                           analysis, and that is the need for a real-time, 2D data presentation package with the
                           ability to present a comprehensive data report providing a list of all peaks found
                           within the 2D space, along with their heights and areas. Ideally, further interpretive
                           data might be provided, such as peak symmetry, resolution of neighbouring peaks,
                           etc. Again, we await the general availability of such software, and perhaps this is the
                           single most important issue to resolve before the methods outlined here become
                           more widely accepted. However, the large data  file sizes obtained with fast data
                           acquisition should not be a problem.



                           4.5.1 COMPARISON OF FIGURES OF MERIT

                           Chromatographically, any procedure will only be of value if the analytical data or
                           figures of merit are acceptable to the analyst. It is necessary to demonstrate the relia-
                           bility of data when using the modulation process. The above points state in very gen-
                           eral terms why modulation gives advantages, but this still requires demonstration. In
                           a recent comprehensive study of analytical figures of merit for the analysis of deriva-
                           tized sterols, we compared the three modes of operation of the LMCS, namely nor-
                           mal capillary GC, comprehensive GC   GC and targeted MDGC (24). This work
                           revealed that improved precision for raw peak areas is achieved with the selective
                           LMCS mode (RSD of  2 %; n   5), over that of normal GC and comprehensive gas
                           chromatography analysis (both with RSD ca.  4 %; n   5). Peaks were generally
                           25–40 times taller for the selective mode when compared with normal GC peaks,
                           based on peak height data, with detection limits of approximately 0.004 mg/l com-
                           pared with 0.1 mg/l for normal capillary gas chromatography and 0.02 mg/l for the
                           comprehensive mode. Note that peak areas are not increased when using the cryo-
                           genic system, although we can measure peak areas down to lower levels. The linear-
                           ity of calibration was best for targeted MDGC (e.g. for the results as obtained by the
                           procedure shown in the lower trace in Figure 4.7), presumably because there is less
                           uncertainty in the peak measurement. Calibration curve linearity for comprehensive
                           GC was similar to normal GC. The better height and area precision for the five repli-
                           cate analyses for selective MDGC analysis may be due to the better confidence of
                           measuring peak response when the peak is much higher than the level of detector
                           noise. The same injection procedure (splitless) was used for all analyses, i.e. with an
                           autoinjector, and hence the reproducibility of injection should have been similar for
                           all of the operational modes.
                              The reason for the differences in comparative figures for the two cryogenic modes
                           may be related to the fact that comprehensive GC gives a series of pulses for each
                           peak, and these must be summed to get the total response, while targeted MDGC
                           gives a single peak for the total component. Normal GC is improved upon by both
                           the other modes. These results indicate that there is every reason to believe that the
                           modulated peak response methods provide reliable and accurate data, at least as
                           good as normal GC, for the analysis of chromatographic peaks.
   106   107   108   109   110   111   112   113   114   115   116