Page 229 - Photodetection and Measurement - Maximizing Performance in Optical Systems
P. 229

Measurand Modulation

            222   Chapter Ten

                                                       Selection
                        Chlorinated                    valve        UV Source
                        sample input


                                                                        Detection system
                                                                        (photometer)
                                               Chlorine-free
                            Dechlorination     sample
                            cartridge                         Absorbance
                                         Waste

                                                                         Time
                        Figure 10.9 A useful technique is to derive the absorbance-free reference liquid
                        from the sample itself, for example via dechlorination.




                          For example, in the detection of trace absorption changes in a more or less
                        pure water sample, we could arrange for sequential measurement of an air ref-
                        erence, a sample of pure water, and the unknown aqueous sample. By measur-
                        ing at the same point in the tubing with a single source-detector system, tubing
                        wall contamination errors can be substantially compensated, while the small
                        differences between sample and solvent can provide a better measure of the
                        small absorption increases.
                          In the measurement of very weak absorptions, it is preferable to use as ref-
                        erence a fluid as similar as possible to the sample itself, so that the detection
                        system can be designed for high amplification of the small differences. This pro-
                        motes configuration 3 of Table 10.1, in which the reference is obtained from the
                        sample stream itself, except purified of the component which is being sought.
                        A typical example might be an optical chlorine disinfectant measurement
                        system, where dechlorination is achieved by passing the water through an acti-
                        vated carbon filter (Fig. 10.9). In practice it is very difficult to remove only
                        the target species without also removing some of the background matrix
                        absorption.
                          Chemical indicators can be used for high sensitivity determinations of a wide
                        variety of chemical species. The reactions are often selective, and the intense
                        colors of some of the reagents give much higher sensitivity than when species
                        absorption is directly measured. Nevertheless, compensation for the absorption
                        of the sample stream is necessary for the best performance, and FIA is a con-
                        venient way to automate this. Verma et al. (1992) describe a system for resid-
                        ual chlorine detection in which the reaction product is a type of azo dye which
                        absorbs at 532nm. This wavelength is much easier to detect, and water absorp-
                        tion is much lower there than at the chorine absorption wavelengths of 234nm
                        and 292nm. The FIA system used performed measurements at a rate of
                                -1
                        110 hour .

                   Downloaded from Digital Engineering Library @ McGraw-Hill (www.digitalengineeringlibrary.com)
                              Copyright © 2004 The McGraw-Hill Companies. All rights reserved.
                               Any use is subject to the Terms of Use as given at the website.
   224   225   226   227   228   229   230   231   232   233   234